Test Parameters and Values
The following parameters and values are available for temporary program-
ming depending on the permanent pacing mode and the selected test mode.
Pacing Mode
1
DDD, ODO, AAI, AOO, OAO, AAT, VVI, VOO,
OVO, VVT
Lower Rate DDD test mode — 30 to 175 ppm
(Lower rate values will be limited by the present Upper
Tracking Rate setting.)
Pacing Rate Single chamber test mode — 30 to 400 ppm
Amplitude 0.5 to 4.0 V (in steps of 0.5 V), 5.0. 7.5 V
Pulse Width 0.03 to 0.15 ms (in steps of 0.03 ms)
0.15 to 1.0 ms (in steps of 0.05 ms)
1.0 to 1.5 ms (in steps of 0.1 ms)
Sensitivity Atrial — 0.18
2
, 0.25, 0.35, 0.5, 0.7, 1.0, 1.4, 2.0, 2.8,
4.0 mV
Ventricular — 1.0, 1.4, 2.0, 2.8, 4.0, 5.6, 8.0, 11.2 mV
1
Temporary test modes are available as follows:
DDD is available as a temporary mode only if the permanent mode is DDD or DDDR.
ODO is available if the pacemaker is a dual chamber model, regardless of the permanent
mode.
If the selected pacemaker is a Prodigy VDD model, temporary modes will be limited to
single chamber ventricular modes.
If the pacemaker is a single chamber model, the test modes will be limited to those
applicable to the paced chamber.
2
The 0.18 mV sensitivity value is not available in Prodigy SR and Prodigy S models.