Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Test Successful
The results of a successfully executed test displays the P-wave or R-wave
amplitude as being between two sensitivity settings. This information can
be used to determine an appropriate permanent sensitivity setting.
To program a permanent Sensitivity setting:
1. Use the [∧
∧∧
∧] or [∨
∨∨
∨] buttons to select a desired pending value. (The
default values displayed are the present permanent values).
2. Position the programming head and press [
PROGRAM
]. Watch for
confirmation of the programming.
Retest Required
Sensitivity Test Value Too Low — Test results that state that the P-wave
or R-wave amplitudes are “greater than” a particular value indicate that
sensing occurred at all of the four sensitivity values tested. Included on the
display is the message, “Rerun test at a higher sensitivity value.”
In this case, you should:
1. Return to the Test Setup screen (select [
RETURN TO SETUP
]).
2. Select a higher test value for sensitivity.
3. Repeat the test starting at step 3.
Sensitivity Test Value Too High — If the test determines that sensing did
not occur at the next setting above the test value, the programmer displays
the test results, but also displays the message, “Rerun test at a lower sensi-
tivity value.”
Determining and Evaluating Parameter Settings 7-19