Specifications

7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Determining a Setting for Sensitivity
The Sensing Test can assist you in determining an appropriate setting
for pacemaker sensitivity. This test measures and reports the present level
of R-wave or P-wave sensing and includes the option for selecting and
programming a permanent sensitivity value based on the test results.
Selecting the Sensing Test Setup Screen
To display the Sensing Test Setup screen, select the [
TESTS
/
ECG
] menu
button, and then select the
SENSING TEST
option as illustrated below.
If an appropriate interrogation has not occurred, a screen message will
direct you to interrogate the pacemaker to display the setup screen. In this
example, the programming head has been positioned over the patient’s
pacemaker to display Marker Channel and EGM telemetry.
1. Select [
TESTS
/
ECG
]
menu button.
2. Select
SENSING TEST
option.
Sensing Test Setup
Screen
Refer to pages 5-6 and
5-7 for instructions on
adjusting the ECG and
telemetry traces.
Determining and Evaluating Parameter Settings 7-15