Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Estimated Time to Replacement
The “Estimated Time to Replacement” is a calculated estimate of the
average time (in months) remaining until pacemaker replacement will be
required. This calculated estimate is based on the pending values selected
for programming as well as data accumulated by the pacemaker since the
previous patient session.
This estimate provides an indication of how different pulse width and
amplitude settings can affect the projected pacemaker longevity.
NOTE: A value for Estimated Time to Replacement will not be displayed if
up-to-date Battery/Lead data is not available. To obtain a value, position the
programming head and select [
INTERROGATE
] as prompted.
Printing the Strength-Duration Threshold Test Results
You can print the strength-duration curve only while it is displayed on the
screen. The graph is not saved for printing later. To print the Strength-
Duration Curve and Threshold Results report:
1. Select the [
PRINT
] Menu button.
2. Select the
CURRENT SCREEN REPORT
option.
You can print the numerical results of the threshold test at any time during
the patient session (see page 7-14). The numerical results are automatically
saved when you exit the threshold test results screen.
Exiting the Strength-Duration Threshold Test Results Screen
If you want to exit the threshold test results screen and conduct another
threshold test, select the [
RETURN TO START
] button.
To exit the Threshold Test function, select the desired menu button from
the top of the screen.
7-10 Using the Programmer