Instruction manual
Measuring Hints
See Figure 5.
• Always use fresh buffers for calibration. Choose buffers that
are no more than 3 pH units apart.
• Check electrode slope daily by performing two-buffer
calibration. Slope should be 92 - 102%.
• Except as noted in Sample Requirements, only use ROSS
™
Internal Filling Solution, Orion 810007, for ROSS Combination
pH and Reference Electrodes. Do not use any filling solution
which may contain silver.
• Remove filling hole cover during measurement to ensure
uniform flow of filling solution.
• Between measurements, rinse the electrodes with distilled
water and then with the next solution to be measure.
• Stir all buffers and samples
• Place a piece of insulating material (e.g. styrofoam or
cardboard) between magnetic stirrer and beaker to prevent
error from transfer of heat to sample. Since ROSS Electrodes
respond faster than conventional electrodes, changes in pH
which result from temperature changes will be noticed.
• Avoid rubbing or wiping electrode bulb, to reduce chance of
error due to polarization.
• After use in especially dirty or viscous samples or when
electrode response becomes sluggish, empty the electrode
completely and hold the junction open under running water.
Empty any water from the electrode and refill with fresh ROSS
Filling Solution. Briefly flush the junction to ensure that the
junction contains fresh filling solution. Maintain a proper level
of filling solution.
Figure 5: measuring Hints
6
Filling hole should
be uncovered
Fresh pH buffer
Stir all samples
and buffers
Filling solution level
must be higher than
sample level
Reference junction
must be immersed
Place insulation
between stirrer
and beaker