Instruction manual
Placement of Removable Sleeve Bulb Guard
The bulb guard can be removed for cleaning or general measurement.
Grasp bulb guard and gently pull. The bulb guard will slide off the
electrode exposing the pH sensing bulb and junction area.
See Figure 4A.
Press on the cap to open junction for easy cleaning.
To replace bulb guard, gently push into place. The bulb guard is
secure without additional turning or fittings. See Figure 4B.
For general measurements the bulb guard can be left in place on the
electrode. However, the bulb guard can contribute to carryover from
solution to solution . When using it, be sure to rinse thoroughly
between standards and samples to prevent carryover. For
measurement in viscous solutions or other solution which may be
retained in the protective guard area, remove the bulb guard for
measurement and calibration. Exposing the pH bulb completely
gives faster response and reduces carryover.
Sure-Flow
®
Junction
The Sure-Flow junction is operable with or without the bulb guard in
place. Simply press down on the cap to open junction and let fill
solution flow through, removing contamination. Removing the bulb
guard and flushing with distilled water aids in the removal of more
difficult deposits.
Figure 4A: Bulb Guard for 8165
Figure 4B: Bulb Guard for 8175
5
O-ring
O-ring