Instruction manual

Sample Requirements
One of the benefits of the ROSS
pH Electrodes is that the filling
solution composition may be changed depending on sample
requirements because of the double junction construction. Do not
use solutions containing silver.
The ROSS pH Electrode Filling Solution, Orion 810007, is 3 M KCl.
For solutions which precipitate in the presence of chloride ion, the
ROSS pH Electrode could be filled with 10% KNO
3
, Orion 900003.
In organic solutions a minimum of 20% water must be present in
the sample. If there is a great deal of drift in organic solutions when
using the ROSS Electrode filled with ROSS Internal Filling Solution
(Orion 810007), try filling the ROSS Electrode with a mixture of
methanol and water saturated with KCl.
The major advantage of the ROSS Sure-Flow
®
Electrode is that the
junction can be cleaned at any time by briefly depressing the cap.
See Figure 3. This can be done whenever necessary. To be sure to
maintain a proper level of filling solution, refill as in step 3 of
Electrode Preparation.
Figure 3: ROSS Sure-Flow Electrode Being Flushed
4