Instruction manual

USING THE ELECTRODE
Electrode Preparation
1. Remove the protective shipping cap from sensing element and
save for storage.
2. Clean any salt deposits from exterior by rinsing with
distilled water.
3. Uncover filling hole by removing plastic sleeve and add ROSS
Internal Filling Solution, Orion 810007, to electrode.
See Figure 2.
To maintain an adequate flow rate, the level of filling solution
must cover the end of coil and be at least one inch above the
sample level when immersed. The filling hole should be open
whenever the electrode is in use.
4. Thoroughly wet the junction by pressing down on the electrode
cap and allowing some of the ROSS Internal Filling Solution to
flow out of the electrode through the junction. Replenish lost
filling solution. See Figure 3.
5. Shake down the electrode (as a clinical thermometer) to
remove air bubbles.
6. Soak combination electrode and/or pH half cell in pH Electrode
Storage Solution, Orion 910001, for one hour. Reference
electrodes do not need to be soaked. If Orion pH Storage
Solution is not available, use 200 mL pH 7 buffer to which
about 1 g KCl has been added, as a
temporary substitute.
7. Connect electrodes to meter.
Figure 2: Filling the ROSS Sure-Flow Electrode
3