Instruction manual
Single Buffer Calibration
1. Ensure that all buffers are at room temperature.
If samples are at varying temperatures, temperature
compensation is recommended. (See meter
instruction manual.)
2. Set up meter according to meter instruction manual.
3. Rinse electrode first in distilled water and then with the
buffer being used for calibration (the buffer should be near the
expected sample pH). Place the electrode in the buffer.
4. Wait for a stable display. Set the meter to the pH value of the
buffer at its measured temperature. See Table 1. (A table of pH
values at various temperatures is supplied on the buffer bottle.)
Proceed to pH Measurement section.
pH Measurement
1. Calibrate the electrode as described in above procedure.
2. Rinse the electrode with distilled water and then with sample.
3. Place the electrode in the sample.
4. When the display is stable, record sample pH.
Electrode Maintenance
1. Inspect the electrode for scratches, cracks, salt crystal
build-up, or membrane/junction deposits.
2. Rinse off any salt build-up with distilled water, and remove any
membrane/junction deposits by holding junction open under
running distilled water.
3. Drain the reference chamber, flush it with fresh ROSS
®
Internal
Filling Solution, Orion 810007, and refill chamber.
Cleaning Electrode
General — Soak in 0.1 M HCl or HNO3 for half an hour, followed
by soaking in pH Storage Solution for at least one hour.
8