Electrical problems (power failure, etc.)
ww
ww
WW
WW
AA
AA
RR
RR
NN
NN
II
II
NN
NN
GG
GG
If a circuit breaker opens frequently, do not use the system. Have
an approved service representative repair the system
Symptom Problem Solution
Mains indicator is not on. The electrical power
cable is not connected.
Connect the power cable.
The inlet circuit breaker
(toggle switch) is open.
Close the circuit breaker
(Figure 5-1).
The power cable is
damaged.
Replace the power cable.
The electrical socket the
power cable connects to
has no power.
Use a different electrical
socket.
An internal fuse is open. Have a service
representative repair the
system.
The electrical outlets do not
have power but the mains
indicator is on.
The main outlet circuit
breaker is open.
Close the circuit breaker.
One electrical outlet does not
have power.
The outlet circuit breaker
is open.
Close the circuit breaker.
A circuit breaker opens
frequently.
Equipment connected to
the outlet(s) uses more
current than the circuit
breaker rating
(Figure 5-1).
Use a different power
supply for some of the
equipment.
The equipment
connected to the outlet
has a short.
Do not use the equipment
until it is repaired.
Tec 6 has no power. Not plugged into outlet. Connect power cable.