Specifications
Aestiva
5-2 1006-0939-000
5-2
About alarms
w WW
WW
AA
AA
RR
RR
NN
NN
II
II
NN
NN
GG
GG
If an alarm occurs, safeguard the patient first, before troubleshooting
or repair procedures.
Two areas on the screen show alarms. The area at the top of the display shows
most alarms. If there are more than 4 alarms at the same time, the lower priority
alarms cycle every two seconds.
During severe malfunctions that prevent mechanical ventilation and/or
monitoring, the area under the waveform shows minimum system messages.
During normal operation, this area shows instructions (push the knob, etc.).
Alarm priority depends on the level of danger to the patient. High priority alarms
require immediate attention.
AB.90.025
Alarms
Priority Alarm tone Alarm silence Note
High 10 tones,
10 second pause,
repeat
120 seconds or
cannot be silenced
Reverse video
Medium 3 tones,
25 second pause,
repeat
120 seconds ---
Low Single tone Tone does not repeat ---