Specifications

4 User Maintenance
1006-0939-000 4-21
4-21
O
2
sensor calibration - 100% O
2
Note: If w appears on the screen, the 21% O
2
calibration must be completed before the
100% O
2
calibration. Mechanical ventilation must be off to perform calibration.
This procedure takes three minutes or less.
Step 1
Push the menu key.
Step 2
Turn the knob to select
Setup/Calibration
(highlight).
Step 3
Push the knob to show the
next screen.