Specifications

Aestiva
4-20 1006-0939-000
4-20
If the calibration fails:
perform the calibration again.
do a 100% O
2
sensor calibration. If this passes, repeat the 21% O
2
calibration.
If the calibration passes, install the O
2
sensor. If necessary, do the 100% O
2
calibration.
Before you use the system, complete the preoperative test procedure. Refer to
"Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.