Specifications

Aestiva
4-18 1006-0939-000
4-18
O
2
sensor calibration - 21% O
2
Note: The 21% O
2
calibration must be done before the 100% O
2
calibration. During O
2
calibration the screen replaces O
2
data with - - -. Mechanical ventilation must be off to
perform calibration.
This procedure takes three minutes or less.
Step 1
Push the menu key.
Step 2
Turn the knob to select
Setup/Calibration (highlight).
Step 3
Push the knob to show the
next screen.