Specifications
4 User Maintenance
1006-0939-000 4-15
4-15
Bellows tests
w WW
WW
AA
AA
RR
RR
NN
NN
II
II
NN
NN
GG
GG
Objects in the breathing system can stop gas flow to the patient. This can
cause injury or death:
• Do not use a test plug that is small enough to fall into the breathing
system.
• Make sure that there are no test plugs or other objects caught in the
breathing system.
w WW
WW
AA
AA
RR
RR
NN
NN
II
II
NN
NN
GG
GG
The bellows assembly test does not replace the preoperative tests. Always
complete the tests in the section Preoperative Tests before you use the
system with a patient.
This test makes sure that all components are correctly assembled. It is not an alternative
to a complete system checkout.
If the bellows operates correctly, install it in the system.
If there is a problem, disassemble the bellows. Look for and replace damaged parts.
Step 1
Hold the bellows assembly
vertical and close the
ports (A and B).
A
B