•To clean the circuit O
2
sensor, wipe it with a damp cloth. Do not put the sensor
in liquid.
• To clean/disinfect metal/or plastic flow sensors, use the flow sensor
cleaning procedure. Do not get the connectors wet.
• Disassemble the bellows before you wash it. If not, it will take a very long time
to dry. Hang the bellows upside down to dry.
• Assemble the bellows before you autoclave. Autoclave the bellows upside
down.
w WW
WW
AA
AA
RR
RR
NN
NN
II
II
NN
NN
GG
GG
Do not use talc, zinc stearate, calcium carbonate, corn starch or
equivalent materials to prevent tackiness. These materials can go
into the patient’s lungs and airways and cause irritation or injury.
w CC
CC
AA
AA
UU
UU
TT
TT
II
II
OO
OO
NN
NN
Do not put the circuit O
2
sensor or flow sensor connector in liquid.
w Do not autoclave the Circuit O
2
sensor or the plastic flow sensors.
w Do not clean the interior surfaces of the flow sensors. Use a damp
cloth
on external surfaces only.
Assemble
The Setup section tells you how to assemble the breathing system.
Inspect all parts for deterioration. Replace them if necessary.
The Preoperative tests in Part 1 (the first half of this manual), tell you how to
test the system for correct operation.
AB.23.088
* Hang the bellows upside down
(extended) to dry. If not, the
convolutions can stick together.
*