Specifications
Aestiva
7-2 1006-0939-000
7-2
Ventilator theory
General The ventilator pneumatics are at the rear of the breathing system.
A precision valve controls gas flow to the patient.
During inspiration, this gas flow closes the exhalation valve and pushes the
bellows down.
During expiration, a small flow pressurizes the exhalation diaphragm to supply
PEEP pressure.
Volume and pressure measurements come from flow sensors in the flow sensor
module. Two tubes from each sensor connect to a transducer that measures
the pressure change across the sensor, which changes with the flow. A third
transducer measures airway pressures at the inspiratory flow sensor.
With circle circuit modules, volume monitoring uses the right (expiratory) flow
sensor. The ventilator uses the other sensor to adjust its output for changes in
fresh gas flow, small leaks, and gas compression upstream of the breathing
circuit. There is no adjustment for compression in the patient circuit. If
necessary, add the compression loss to the tidal volume setting (volume
control mode). The average volume changes from compression in the breathing
circuit is small (0.5 to 1.25 mL/cmH
2
O).
For better precision:
•When the fresh gas mixture includes Heliox, use the Heliox mode (Ventilation
setup menu). Heliox, used on some ANSI Models of the Aestiva, changes the
data collected by the flow sensors. When Heliox mode is selected, the
ventilator adjusts the flow sensor data to correct for these changes.
•A small quantity of gas bleeds through a resistor to help keep the pressure on
the exhalation valve constant. At high airway pressures, this can cause a
slight hiss during inspiration.
ww
ww
WW
WW
AA
AA
RR
RR
NN
NN
II
II
NN
NN
GG
GG
Do not try to silence the pneumatic resistor. If it is blocked, the
ventilator can malfunction and cause patient injury.
Always connect the expiratory flow sensor. If it is not connected,
the patient disconnect alarm can not operate correctly.










