UM-0085-B09  DT80 Range User Manual  Page 401 
RG 
  Error Number – Description 
•  Cause/Action 
Error Category 
Syntax  Operation  Memory  Reading  Hardware 
  E160 – Invalid password – detailed msg 
•  The password is invalid 
  E161 – User accounts limit – 
detailed msg 
•  Cannot create a user account as soon as maximum 
number of account in the user’s database reached. 
 
  E162 – User authentication required – 
  detailed msg 
•  User authentication is required 
 
  E163 – Function not implemented – 
detailed 
  msg 
•  Function is not implemented yet 
 
  E164 – User not logged in – detailed msg 
•  User not logged in 
 
  E165 – Authentication disabled – 
detailed msg 
•  No security related commands are allowed when 
authentication is disabled 
 
  E166 – No users configured – detailed msg 
•  No one user account is configured 
 
  E167 – User change role – detailed msg 
•  Changing role for the current user is not allowed 
 
  E168 – No users configured – detailed msg 
•  Operation failed as authentication is enabled 
 
Table 24: DT80 Error Messages 
Data Errors 
Errors may also occur even though the DT80's measurement system is operating normally – for example if an analog 
input is out of range or a connected sensor does not perform correctly. In some cases, the DT80 returns an error 
message (see "Reading" error category in the above table), but mostly there is no message returned. Instead, the DT80 
flags the logged and/or returned data value as invalid. 
Each logged reading has an associated "data state", which identifies whether that particular reading is valid or invalid. 
The following data states are possible: 
Logged data 
state 
Value when unloaded 
CSV/free format 
Fixed format  Description 
OK  measured value  measured value  No error 
Overrange 
OverRange  99999.9 
input voltage exceeds max positive input voltage 
input voltage exceeds positive linearization limit for sensor 
input frequency too high (
F
 channel type) 
Underrange 
UnderRange  -99999.9 
input voltage exceeds max negative input voltage 
input voltage exceeds negative linearization limit for sensor 
input frequency too low (
F
 channel type) 
Not Yet Set 
NotYetSet  -9e9 
DT80 analog hardware fault 
insufficient samples have been taken to calculate a statistical 
value 
could not read a valid value from a serial device 
Reference Error 
RefError  8e9 
Thermocouples and some bridges require a separate reference 
channel measurement as part of the measurement process. 
This error indicates that although this channel's raw 
measurement was OK, the associated reference channel 
measurement was not. 
Calculation Error 
Error 
-8e9 
Maths error such as division by zero 
Table 25: DT80 Data Error Value 