Instructions
UM-0085-B09  DT80 Range User Manual  Page 308 
RG 
4–20mA Current Loops 
Many different sensor types provide a 4-20mA current output, where the current is proportional to the quantity being 
measured. 
The channel type for a 4-20mA current loop measurement is 
L; channel options are as for Current measurements; see 
Current (P306). 
To read a current loop sensor, the DT80 first measures the current. Any of the Current wiring configurations may be 
used. The measured current is then scaled so that the channel returns 0% for 4mA and 100% for 20mA. 
For example, if a current loop sensor is connected as per C3 – Independent Current Input using Internal Shunt (P307) 
then it would be read using 1#L. 
A span is often applied to this value so that the final reported value is in the proper engineering units for the quantity 
being measured. 
For example, if a pressure sensor with 4-20mA output operates over a range of 100-500kPa then 
BEGIN S1=100,500"kPa" RA2S 4#L(S1) END 
will return a pressure reading in kPa every 2s. The loop would in this case be connected across the 4# and AGND/EXT# 
terminals, making use of the internal shunt resistor. 
One advantage of a current loop system is that a cable fault can be readily detected. An open circuit loop will give a 
negative reading, normally -25% (0mA), which can then be detected in an 
ALARM statement. 
  Multiple Devices in One Loop 
Multiple measurement devices can be connected in series in the one current loop. For example, a current loop might 
include both a DT80 and a 4-20mA digital panel display module, in series. (There can only be one 4-20mA sensor
 in a 
given loop.)  
In these situations care needs to be taken to ensure that the DT80's common mode voltage limits are not exceeded, 
especially if the DT80 is placed "above" the other devices in the loop. The DT80's input attenuators may be enabled 
using the 
A channel option, which will increase the common mode limit to ±35V, relative to AGND/EXT# 
Note: If there are multiple devices in the loop then the loop should be continuously powered. This implies the use of an external shunt 
(C1 – Shared Terminal Current Inputs using External Shunts (P306
), as the DT80 normally disconnects the EXT# terminal between 
measurements. 
Resistance 
Resistance (degree of impediment to current flow) is measured by passing a known excitation current through the 
resistance and measuring the voltage across it. From Ohm's law, Resistance = Voltage / Current. 
The R channel type returns the resistance value in ohms (Ω). 
The DT80 Series 2, 3 incorporates two and Series 4 has a three precision current sources. When measuring a 
resistance, one or other of these sources is switched through to the * (Excite) terminal, which is then connected to the 
unknown resistance. (On Series 3 models, excitation can also be directed to the + or – terminals.) 
By default, an accurately known current of approximately 200μA is generated by the DT80. This allows measurement of 
resistances up to about 10kΩ. For low resistances (up to 700Ω), the 
II channel option can be used to select a higher 
excitation current (2.5mA), which will allow a more accurate reading. In Series 4 an additional current source 
III with 
current of approximately 2uA has been added to increase datataker capabilities to measure resistances up to 1MΩ. 
One problem which occurs when measuring low resistance values is the fact that the wires used to connect the unknown 
resistance to the logger also have resistance. As shown in the wiring diagrams below, there are various ways to 
overcome this. These involve the use of separate wires to carry the excitation current and to sense the voltage. 
For example, the channel 
1R(4W) will perform a four-wire measurement. Excitation current flows out the * terminal, 
through the unknown R, and returns to the # terminal. Two separate sense wires connect the + and – terminals to the 
unknown R. Because the DT80 has a very high input impedance (>10MΩ), negligible current will flow in these sense 
wires, resulting in a negligible voltage drop. The measured voltage will therefore be the voltage across the unknown R 
only, and will not include any voltage drop in the current-carrying excitation cables. 
The downside of a four-wire measurement is that you need, well, four wires. The three and two wire configurations 
provide for reduced cable cost, at the expense of accuracy. 










