Instructions
UM-0085-B09  DT80 Range User Manual  Page 384 
RG 
System 
Display and Keypad  
•  Type: LCD, 2 line by 16 characters, backlight. 
•  Display Functions: channel data, alarms, system status. 
•  Keypad: 6 keys for scrolling and function execution. 
•  Status LEDs: 4 (Sample, Disk, Attention, Power) 
Firmware Upgrade 
•  RS232, Ethernet (FTP), USB or USB disk. 
Real Time Clock 
•  Resolution: 0.2ms 
•  Accuracy: ±1 min/year (0°C to 40°C), ±4 min/year (-40°C to 70°C) 
Power Supply 
•  External power input: 10 to 30Vdc 
•  Internal 6V lead acid battery: DT80/81/82I: 1.2Ahr, DT85/85G: 4.0Ah, other models: none 
•  Power output for charging external lead acid battery (some models only) 
Power Consumption 
•  Peak Power: 12W (12Vdc 1A) 
•  DT80/81/85 approximate average power consumption (10 analog channels): 
Schedule Rate 
Ext 12V power consumption 
1.2Ah battery life 
4.0Ah battery life 
1 sec 
1400 mW 
6 hours 
1 day 
5 sec 
500 mW 
1 day 
3 days 
1 minute 
100 mW 
10 days 
1 month 
1 hour 
60 mW 
3 months 
9 months 
•  DT82E/80L/85L approximate average power consumption (6 analog channels): 
Schedule Rate 
Ext 12V power consumption 
1 sec 
750 mW 
5 sec 
300 mW 
1 minute 
40 mW 
1 hour 
10 mW 
Physical and Environment 
•  Construction: Powder coated zinc and anodized aluminium. 
•  Dimensions: DT80/81: 180 x 137 x 65mm, DT85: 300 x 137 x 65mm 
•  Weight: DT80/81: 1.5kg, DT85: 2.5kg 
•  Temperature range: –40°C to 70°C (integrated modem: –30°C to 70°C) * 
•  Humidity: 85% RH, non-condensing 
* reduced battery life and LCD operation outside range –15°C to 50°C 
Accessories Included 
•  Resource CD: includes software, video training, and user manual. 
•  USB communications cable (DT80/81/85) 
•  Micromag multiband antenna with 2m cable (DT80LM/82EM/85LM/85GLM) 
•  Ethernet crossover cable (DT82) 
•  Line adaptor: 110/240Vac to 15Vdc, 800mA 