Instructions
UM-0085-B09  DT80 Range User Manual  Page 55 
RG 
  Example 
The schedule 
RDX 1..3TK 
samples analog channels 1 to 3 as type K thermocouples (
1..3TK) whenever the DT80 receives an XD poll command 
(that is, whenever it receives the character sequence 
XD) either from a connected computer, or by means of an alarm 
action from within the DT80. 
  Using Poll Commands with Standard Report Schedules 
Figure 12: Polling a time interval schedule 
A report schedule defined with a time or event trigger can also be polled by its appropriate poll command at any time. For 
example, the report schedule 
RC5M 1V 2V 3V 
normally runs every 5 minutes (5M), but it can also be run at any time by an 
XC poll command (from the host computer 
or an alarm). 
For schedules that have a long interval, this is useful for checking that a sensor is functioning. 
Continuous Trigger 
Figure 13: Continuous schedule 
Report schedules that run continuously can be created. A continuous schedule will start scanning as soon as job is 
loaded onto the DT80, and run until it is stopped (by sending a halt command or resetting the DT80, for example). 
Define a continuous schedule simply by omitting the trigger from a report schedule. 
  Example — Continuous Schedule 
Sending 
RA 1TK 2R(3W) 3TT 
causes the DT80 to scan the specified channels continuously. 
Note: The X schedule does not support continuous trigger. The schedule definition RX is equivalent to RXX, i.e. it defines a polled 
schedule, not a continuous schedule. 
Trigger While 
Figure 14: Schedule "while" condition 
A report schedule’s trigger can be enabled or disabled by an external condition. This is called the While condition — that 
is, trigger only while
 the external or internal condition is true. 
The While condition can be either 
•  states on one or more of the DT80’s digital input channels (n
DS), or 
•  internal conditions specified to the DT80 as states of channel variables: 










