Instructions
UM-0085-B09  DT80 Range User Manual  Page 397 
RG 
Error Messages 
Standard Messages 
The DT80 returns a message when it detects an error in a command, or an operational difficulty. The form of the error 
report is controlled by the /U switch. The default is the verbose form shown in the table below. If the switch is set to /u 
the error message is reduced to an error number (e.g. E3). (Note this Switch also reduces the verbosity of other returned 
data). 
Error messages can be switched off by the 
/m switch. The default is for errors to be reported (/M). During an unload 
operation, error reporting is disabled until the unload is complete. If an error is detected during job entry (i.e. between 
BEGIN and END) then the remainder of the job is ignored. 
The table below lists all of the DT80 errors, along with an explanation of their likely cause and/or correction. 
  Error Number – Description 
• 
Cause/Action 
Error Category 
Syntax 
Operation 
Memory 
Reading 
Hardware 
  E1 – Time set error 
•  Must be in format defined by P39 and P40 
•  Illegal separator or non-digits entered 
 
  E2 – Command line too long 
•  Command too long (maximum 1023 characters) 
 
  E3 – Channel option error 
•  Illegal channel option used 
•  Incompatible options used 
•  Option invalid for channel type 
 
  E7 – Day set error  
•  Illegal day number entered 
 
  E8 – Parameter read/set error  
•  Parameter indEX out of range 
•  Parameter value out of range 
 
  E9 – Switch error  
•  Illegal switch command character 
 
  E10 – Command error  
•  Unrecognised keyword 
 
  E12 – Channel list error 
•  Channel number outside the legal range 
•  Incomplete channel sequence 
•  Invalid channel type 
•  Polynomials or spans indEX out of range 
 
  E18 – STATUS command error 
•  STATUSn outside the range 1 to 14 
 
  E23 – Scan schedule error 
•  Schedule ID not A-K, X or S 
•  Scan time interval too large 
•  Scan interval type invalid 
•  Event or counter channels invalid 
 
  E25 – Channel table full 
•  Internal acquisition and alarm table filled (maximum 
1000 entries). Additional channels cannot be declared 
 
  E29 – Poly/span declaration error 
•  Polynomial or span indEX out of range (1 to 50) 
•  Individual terms not separated by a comma 
•  Range of terms outside 1.0e–18 to 1.0e18 
 
  E30 – Analog measurement fault 
•  Faulty circuit board, circuit board connector, or circuit 
board power supply 
•  If fault persists after a hard reset contact your 
dataTaker representative. 
 
  E32 – Job not found 
•  The named job cannot be found. 
 










