Instructions
UM-0085-B09  DT80 Range User Manual  Page 374 
RG 
Getting Optimal Speed from Your DT80 
In applications where it is important to sample as rapidly as possible, the following guidelines may assist: 
•  Switch off data return and logging for channels you are not interested in. 
•  Set 
P21=1 so that the analog power supply is always on. This prevents the 50ms warm-up delay which can 
occur if there is any gap between schedule execution. 
•  Similarly, if any CEM20 units are in use then set P28=1 so that 12V output for powering the CEM is always on, 
•  Set P11 to a higher value, which will reduce the time over which an analog acquisition is integrated. For 
example, if P11 is changed from 50 to 500 then the sample time will be reduced from 20ms to 2ms. Be aware 
that this will cause a degradation in the DT80's ability to reject noise. 
•  Reduce the channel's settling time, using the MDn channel option. The default is 10ms. It is not recommended 
to reduce this below 5ms because the DT80's relays need about this long to switch before a measurement can 
commence. 
•  Disable automatic calibrations using 
/k. Be aware that readings will now be subject to drift with temperature. 
•  Set 
P62=1, which will leave the DT80's relays set when the schedule completes (normally they are set back to 
a quiescent state at the end of a schedule). If you are measuring a single voltage input then you will now be 
able to do repeated samples without changing the relay settings at all, which would then allow you to reduce the 
settling time to zero (i.e. 
MD0) 










