Instructions
UM-0085-B09  DT80 Range User Manual  Page 338 
RG 
  Schedule Triggers 
Counter channels can be configured to trigger a schedule when the counter reaches its specified wrap value (at which 
point it resets to 0). See Trigger on External Event (P53) for more details. For example, 
RA2C(10) 1TK 
will measure a temperature on every 10th pulse received on digital input 2D. 
  Presetting Counters 
The count value for a digital input channel can be preset using an expression, e.g. 
RA1M 8C=1000 RB2S 8C 
will reset the counter to 1000 once per minute. So if a 1Hz signal is now applied to input 8D you would expect the values 
returned every 2s for channel 8C to follow a sequence similar to: 
  1000, 1002, 1004 ... 1056, 1058, 1000, 1002 ... 
  Setting Counter Wrap Value 
A counter’s wrap value (channel factor) is applied when the channel is defined (i.e. when the job is entered), not when it 
is evaluated
. Also, setting the wrap value has the side effect of resetting the count value to zero. This implies that: 
•  a particular counter’s wrap value need only be specified once in the job. It does not need to be specified every 
time the counter is evaluated. 
•  If querying a counter using the immediate schedule (e.g. by periodically typing "
1C"), do not specify a wrap 
value each time. Each time you evaluate an immediate channel you are also defining it, so the counter value will 
always be returned as zero if you specify a wrap value each time. 
Counters – High Speed  
The DT80 provides four dedicated counter inputs, labeled 1C-4C (7 counters 1C-7C on DT85/85L Series 3). These 
provide for high speed pulse counting (100kHz typical) and can also function as phase encoder inputs. 
The HSC (high speed counter) channel type is used to access these counters. For example, 2HSC will return the number 
of positive transitions seen on the 2C counter input. 
Figure 162: High Speed Counter Channels 
This above figure shows a simplified circuit diagram for the DT80’s hardware counter inputs. As can be seen, the 
channels can be divided into pairs of inputs, 1C-2C (counter channels 1HSC and 2HSC) and 3C-4C (3HSC and 4HSC); 
also 5C-6C on DT85/ 85L Series 3. Each pair can be used as either: 
•  two independent counter inputs, for pulse counting, or 
•  a single phase encoder (quadrature) input, for use with position sensors that provide phase encoded outputs 
("A" and "B"). See Phase Encoders (P340) 
Note: on the DT81, only the 3C-4C inputs can be used with a phase encoder. 