UM-0085-B09  DT80 Range User Manual  Page 119 
RG 
The following "screens" will then be available. These can be scrolled through using the Up and Down arrows on the 
keypad. Pressing Down 
Display screen 
Comments 
 DT80 V9.00 
 MYJOB 
The default "sign-on" screen indicates the DT80's firmware version number and the name of the 
currently loaded job (No current job is displayed if there isn't one) 
Date: 23/10/2009 
Time: 16:44:02 
Current date and time (format can be changed using P31 and P39) 
Battery: 90% ↓ 
-290mA 6.2V 
Internal battery status. This shows the approximate battery charge as a percentage, a charge (↑) or 
discharge (↓) indicator, battery current (negative=discharging) and the battery terminal voltage. NC 
is displayed if the internal battery is not connected. 
This screen is not present on models with no internal battery charger, e.g. DT82E 
Telstra 3■□□□ 
203.112.100.109 
Internal modem status. While the modem is connected the display shows the mobile carrier name, 
network type (e.g. '3' for 3G), signal strength and IP address. If the modem is not connected then 
modem status is shown e.g. Modem is off. 
This screen is only present on models with an integrated modem (DT8xM) 
Eth IP: Auto 
192.168.11.160 
Ethernet port mode (Automatic, Manual or Disabled) and current IP address 
 Boiler Temp 
 97 °C 
First user channel (user defined channel name) 
 Channel 2LM35 
 17.9 
°
C 
Second user channel (default channel name) 
 Alarm4 OFF 
 1356.3 mV 
Alarm #4 state 
 Oven Over Temp 
 117.2 °C 
Alarm #7 state (alarm text replaces channel name when alarm is active) 
Note: channel 1CV is not displayed because it is defined as a working (W) channel. Working channels are neither logged, returned nor 
displayed. 
Bar Graph 
The channel value can be shown as a bar graph instead of a numeric value by using the BG channel option. The BG 
option allows the values to be set that represent the left and right side of the graph scale. The channel label can be used 
to set the graph scale labels. For example: 
Display screen 
Comments 
E--Fuel Level--F 
███████ 
4V("E--Fuel Level--F",BG10:900) 
displays zero scale (no bars) if measured voltage < 10mV; displays full scale (16 bars) if voltage > 
900mV 
Controlling what is shown on the display 
All defined channels and alarms will be shown on the display, except for: 
•  channels which specify the 
ND (no display) channel option 
•  working channels (
W channel option) 
•  un-numbered alarms 
•  alarms where the channel in the alarm condition is a no-display (
ND) or working (W) channel 
For example: 
4TT 'displayed 
  3CV(W)=3CV+1  'not displayed
  ALARM1(2CV>10)"Quokka" 'displayed 
  ALARM(2CV>20)"Platypus" 'not displayed 
  ALARM2(2CV(ND)>30)"Echidna" 'not displayed 