Instructions
UM-0085-B09  DT80 Range User Manual  Page 330 
RG 
  Example 
The following example shows how the DT80 might be used to calculate a strain value. It will most likely need to be 
adapted to suit the type of sensor and the application. In this case a 5-wire sensor is used, with the manufacturer 
supplied parameters as shown. 
BEGIN 
   'Manufacturer data 
   'b = 68.22 ohm 
   'a = 4.86 degC/ohm 
   'G = 5.8962 ue/0.01% 
   'Z = 1.0137 
   'C = 10.56 ue/degC 
   RA12H 
   1R(ES5,4W)  'measure R1, take 5 extra samples to reduce noise 
   2R(ES5,4W)  'measure R2, take 5 extra samples to reduce noise   
   'calculate temperature if required 
   CALC("Temp~degC")=((&1R+&2R)-68.22)*4.86  
   'calculate strain 
   CALC("Strain~ue")= 58962*(&1R/&2R-1.0137)-10.56*Temp  
  END 
So in this case if the measured resistances were R1=36.44 and R2=35.56 then the calculated temperature would be 
18.4°C and the strain +457 µε. 
Analog Logic State Inputs 
For reading logic states, the DT80's digital inputs (see Digital Channels (P331)) are the best option. If there are 
insufficient digital inputs then the analog inputs can also function as digital inputs, using the 
AS channel type. 
The AS channel type detects an input voltage relative to a threshold: 
•  When the input is above the threshold, 1 is returned 
•  When the input is below the threshold, 0 is returned 
Any of the Voltage wiring configurations may be used. 
The default threshold is 2500mV, but the channel factor can be used to set any value in mV. 
Input attenuators are enabled by default for the AS channel type, allowing input voltages up to 30V. 
For example: 
1AS(1750) 
configures analog channel 1 as an analog state input with a threshold of 1.75V. 










