Instructions
UM-0085-B09  DT80 Range User Manual  Page 311 
RG 
R6 – High Resistance Input with Parallel Resistor 
Note:  This configuration fits to Series 3 or older as the resistance measurements are limited to a maximum of about 10kΩ. Measuring 
resistance above 1MΩ in Series4 has not been tested 
Resistance measurements are limited to a maximum of about 10kΩ. This can be extended by wiring a known resistor in 
parallel with the resistance being measured. This will, however, reduce the resolution of low resistance measurements.  
Figure 138: Wiring for 4-wire resistance input, using a parallel resistor 
To measure 
Use the command 
R 
  1R(4W,W) 
 CALC("R~ohm")=(Rp*&1R)/(Rp-&1R) 
As shown above, we first read the combined resistance, and then calculate the value of R using an expression that 
references the combined resistance measurement (&1R). Rp represents the value of the parallel resistor in ohms. As 
well as the 4-wire configuration shown here, a parallel resistor can also be used with a 3-wire or 2-wire resistance 
measurement. In all cases, the parallel resistor (Rp) should be located near the sensor
 (R), as shown above, so that the 
lead resistances can be correctly compensated for. 
If it is not practical to locate the resistor near the sensor then it can be located at the logger end of the cable. In this 
configuration the best accuracy will be obtained by connecting the sense inputs (+ and -) across Rp (if its resistance is 
significantly less than R). If Rp is greater than R then the sense inputs should instead be connected across R, although 
in this case the effect of cable resistance is likely to be negligible, given that both R and Rp are high resistances. 
  Calculating Parallel Resistor Value 
The required value of the parallel resistor Rp is given by: 
=
10000 
10000
where R
max
 is the maximum resistance required to be measured. For example, to measure up to 100kΩ a parallel resistor 
of about 10kΩ would be suitable. 
Bridges 
Because of its sensitivity, the Wheatstone bridge circuit is commonly used for the measurement of small changes in 
electrical resistance
. Applications include load cells, pressure sensors and strain gauges. 
Figure 139: Wheatstone bridge 
Bridges are designed such that under quiescent conditions the ratios R1/R4 and R2/R3 are equal, resulting in a zero 
output voltage, V
out
. A small change to one or the resistances will then cause a corresponding change to V
out
, which can 
then be measured accurately using the DT80’s sensitive 30mV range. 
When one of the four resistors in a bridge is active (that is, sensitive to the quantity being measured) the circuit is called 
a quarter bridge, and the remaining three resistors are called bridge completion resistors. Similarly, half and full 
bridges imply two and four active gauges. All completion resistors should be close-tolerance precision resistors. 
The DT80 returns all bridge measurements in a ratio metric form with units of parts per million (ppm): 
=
10
where: 
•  V
out
 is the measured bridge output voltage 
•  V
ex
 is the excitation voltage 
For a bridge measurement to be accurate, both of these voltages must be known accurately, and any lead or connector 
resistances must be compensated for. 










