Instructions
UM-0085-B09  DT80 Range User Manual  Page 275 
RG 
Diagnostic Commands 
TEST Command 
The TEST command causes the DT80 to perform a self-calibration, then run a series of self tests. Test results that are 
out of range are flagged with a FAIL message. 
This command can also be issued via the DT80 web interface . 
A typical test report is shown below. 
TEST 
TEST report generated at 2008/02/13,16:05:09 
 0  dataTaker 80 Version 7.02.0002 Flash 2007/12/21 17:09:56 
29  Product: DT80-2 
 1  Serial Number: 080043 
 2  VEXT 11.0 V 
 3  VBAT (6V) 6.74 V PASS 
 4  IBAT +13 mA PASS 
 5  VSYS 6.71 V PASS 
 6  VLITH (3.6V) 3.68 V PASS 
 7  VDD (3.3V) 3.25 V PASS 
 8  VANA (3.8V) 3.65 V PASS 
 9  VRELAY (4.5V) 4.20 V PASS 
10  VREF (2.5V) 2509.72 mV PASS 
11  Ics I 0.21341 mA PASS 
12  Ics II 2.5755 mA PASS 
13  Vos diff -0.2 uV PASS 
14  Vos 3W I 345.7 uV PASS 
24  Vos 3W II -58.7 uV PASS 
15  Vos shunt -1.2 uV PASS 
16  Vos + 2.4 uV PASS 
17  Vos - 2.9 uV PASS 
18  Vos * -24.3 uV PASS 
19  Vos # 57.6 uV PASS 
25  Vos diff atten -0.3 uV PASS 
26  Vos + atten 64.4 uV PASS 
27  Vos - atten 67.7 uV PASS 
28  Vos * atten 37.2 uV PASS 
20  Term. factor 1.00470 PASS 
21  Shunt (100R) 99.681 Ohm PASS 
22  CMRR 137.3 dB PASS 
23  Overall health PASS 
You can also request just one line of the report using 
TESTn, where n is the line number (0-29, as indicated above). For 
example, the following will return the firmware version number: 
  TEST0 
  dataTaker 80 Version 7.02.0002 Flash 2007/12/21 17:09:56 
Some other useful test report parameters are described below: 
•  Product – this indicates the product type and series, for example DT80-2 indicates a DT80 Series 2. 
•  Serial number – this should match the number on the label affixed to the outside of the unit. 
•  VEXT – the voltage at the external power input. 
•  VBAT – internal battery terminal voltage. 
•  IBAT – internal battery current: positive when battery is charging, negative when discharging. 
•  VLITH – internal memory backup battery terminal voltage. 
The other parameters listed in the test report are generally only useful to dataTaker service personnel. A description of 
these entries is beyond the scope of this manual. 