UM-0085-B09  DT80 Range User Manual  Page 274 
RG 
Safe Mode 
If there is an error in the DT80's profile settings or the job that is automatically loaded following hard reset then you may 
find that you are unable to communicate with the logger. For example – the startup profile specifies the wrong RS232 
parameters (and you're not sure what they are), or the job specifies an immediate 
1SERIAL channel with a very long 
timeout (and the serial channel is not connected to anything). 
Safe Mode provides a mechanism to regain control and fix the problem. To select safe mode you need to perform a 
"triple-push" reset: 
1.  Press the manual reset button 
2.  Wait approximately 3 seconds, then press the reset button again. 
3.  Wait approximately 3 seconds, then press the reset button a third time. 
4.  Verify that DT80 restarted / Safe mode is displayed on the LCD. 
This will have the same effect as a hard reset, except that: 
•  all profile settings are ignored – factory default settings are used (for example, the host port parameters will be 
set to 57600,8,N,1,SWFC) 
•  no job is loaded 
•  any queued unload or alarm messages will not be attempted 
Note: the profile settings are not cleared – they are simply ignored for this session only. If you now do a normal hard reset they will once 
again be loaded. 
Factory Settings 
To restore the DT80 to factory default settings, send the FACTORYDEFAULTS command, which will reset all profile 
settings to their factory defaults, then perform a hard reset. 
Note: this command does not delete any files from the internal file system. The DELALLJOBS command may be used to remove all 
jobs and all logged data. 