UM-0085-B09  DT80 Range User Manual  Page 294 
RG 
The following tables show the equivalent data for the "low power" DT80 models 
Scan rate 
0 analog channels 
5 analog channels 
30 analog channels 
300 analog channels 
continuous 
1300 mW 
2020 mW 
2000 mW 
2000 mW 
1 sec 
370 mW 
720 mW 
2000 mW 
2000 mW 
5 sec  200 mW  280 mW  630 mW  2000 mW 
10 sec 
100 mW 
150 mW 
320 mW 
2000 mW 
1 min  25 mW  35 mW  65 mW  380 mW 
10 min 
12 mW 
12 mW 
15 mW 
50 mW 
1 hour 
10 mW 
10 mW 
11 mW 
17 mW 
10 hour and above 
10 mW 
10 mW 
10 mW 
10 mW 
Table 14: DT82E/ 80L/ 80GL/ 85L/ 85GL core hardware average power consumption (external 12V power) 
Scan rate 
0 analog channels 
5 analog channels 
30 analog channels 
300 analog channels 
continuous 
1400 mW 
2300 mW 
2300 mW 
2300 mW 
1 sec  420 mW  820 mW  2300 mW  2300 mW 
5 sec 
220 mW 
320 mW 
720 mW 
2300 mW 
10 sec  120 mW  170 mW  370 mW  2300 mW 
1 min 
35 mW 
45 mW 
80 mW 
440 mW 
10 min 
21 mW 
22 mW 
26 mW 
65 mW 
1 hour  20 mW  20 mW  21 mW  27 mW 
10 hour and above 
20 mW 
20 mW 
20 mW 
20 mW 
Table 15: DT82E/ 80L/ 80GL/ 85L/ 85GL core hardware average power consumption (external 24V power) 
Note that: 
•  At slow scan rates the logger spends nearly all of its time asleep, so the average power consumption is 
determined solely by the core hardware's sleep mode power consumption, which is fixed. 
•  If sleep is disabled, then for slow scan rates the core hardware power consumption will approach the "idle" 
power state value (300/ 500/ 700 mW for 6/ 12/ 24 V supply, or 350/ 400mW for 12/ 24V supply for low power 
models) 
•  For a continuous schedule, the "30 analog channels" case actually uses slightly less power than the "5 analog 
channels" case. This is because a greater proportion of the time is spent with the processor idle, waiting for the 
analog sampling to complete. 
•  These tables assume that the schedule contains only analog channels. Digital channels can usually be 
disregarded, but if there are a significant number of calculations being done then the average power 
consumption will increase. If any very time consuming operations are performed (e.g. reading SDI-12 sensors) 
then these tables are no longer applicable. Refer to Example 3 (P296) for more information. 
•  A measurement made using a CEM20 channel takes approximately twice as long as a DT80 channel. Thus if 
15 CEM20 channels are sampled, you should use the values in the "30 analog channels" column above. See 
Example 4 (P296). 
For example, if 10 analog channels are being sampled once per minute then interpolating from Table 12 the core 
hardware will draw an average of approximately 100 mW from a 12V external supply (although from Table 10 the peak 
demand may be up to 1800 mW). 