Specifications
HYDROLOGICAL SERVICES
DL3 100-027 IMP
Issue 2: 15/11/2005
On-line Help
PASSWD?
Password: Four character passwd to allow user access
to alter parameters or change existing passwd
NOTE: Upper/lower case sensitive
Syntax: PASSWD=****<CR> To attempt access, or change password if access
already accepted
The reset command performs a hardware reset, but does not affect the time, date, parameters or
logged data.
On-line Help
RESET?
Reset: Performs a reset of the logger. Note this command
is equiv to switching logger off-on again
Syntax: RESET<CR>
On-line Help
VER?
Software Version:
Syntax: VER<CR> Displays the software revision and date
4.1.10Rain Parameters (RCHID, RI)
The "Rainfall Channel ID" (RCHID) is a 7 digit user definable number that should be set to uniquely
set to identify the rainfall.
On-line Help
RCHID?
Rain Channel ID: Fixed length 7 digit numeric field
Syntax: RCHID<CR> Displays rain ch ID
RCHID=0012345<CR> Sets RCHID to 0012345
The "Rainfall Increment" (RI) parameter defines the bucket capacity of the specific TBRG used as
well as the number of tips that defines one rainfall event - this determines the resolution and the rate
at which flash memory is used.
On-line Help
RI?
Rain Increment: Defines water volume that constitutes recorded
rainfall event. Bucket capacity in inch x recording mult
Syntax: RI<CR> Displays the rainfall inc parameter
RI=0.01/02<CR> Defines bucket size as 0.01
capacity & a mult
of 2 => 0.02
of rain is one event