PowerScan® 7000 SRI Product Reference Guide
Datalogic Scanning, Inc. 959 Terry Street Eugene, Oregon 97402 Telephone: (541) 683-5700 Fax: (541) 345-7140 An Unpublished Work - All rights reserved. No part of the contents of this documentation or the procedures described therein may be reproduced or transmitted in any form or by any means without prior written permission of Datalogic Scanning, Inc. or its subsidiaries or affiliates ("Datalogic" or “Datalogic Scanning”).
Table of Contents About This Manual ............................................................................ 1-1 Manual Conventions .................................................................... 1-2 Connecting the Scanner ..................................................................... 1-2 Programming ............................................................................. 1-4 Resetting the Standard Product Defaults .............................................. 1-4 Double Read Timeout .
Bar/Space Polarity ............................................................... 3-34 Wand Idle State .................................................................. 3-34 Signal Speed ....................................................................... 3-36 Transmit Leading Noise ........................................................ 3-37 Transmit Trailing Noise ......................................................... 3-38 Symbology Conversion .....................................................
RSS Expanded 2D Component .................................................... 5-31 RSS Limited ................................................................................... 5-32 Disable/Enable RSS Limited ....................................................... 5-32 UCC/EAN 128 Emulation ............................................................ 5-32 Minimum Reads ........................................................................ 5-33 RSS Limited 2D Component ..................................
Code 11 ......................................................................................... 5-77 Disable/Enable Code 11 ............................................................. 5-77 Number of Check Characters ...................................................... 5-78 Check Character Transmit .......................................................... 5-78 Length Control ..........................................................................
Chapter 1 Getting Started The PowerScan® 7000 SRI (Standard Range Imager) marks a new performance level for handheld scanners. It delivers aggressive read rates and depths of field on 1D and stacked codes. This aggressiveness applies even in challenging reading environments where low lighting conditions and poor quality might make it difficult to read bar codes. You can rest assured your investment will continue to supply years of use by reading any bar codes you require, now or in the future.
Getting Started Manual Conventions The symbols listed below are used in this manual to notify the reader of key issues or procedures that must be observed when using the scanner: Notes contain information necessary for properly diagnosing, repairing and operating the scanner. NOTE The CAUTION symbol advises you of actions that could damage equipment or property. CAUTION Connecting the Scanner Use the appropriate instructions below to connect the scanner to the terminal, PC or other host device.
Connecting the Scanner Figure 1. RS-232 Serial Connection Cable Connector Retainer Boss Captive Screws USB Connection — Connect the scanner to a USB port on the terminal/PC using the correct USB cable for the interface type you ordered. Reference Figure 2. B Figure 2. USB Connection or... nd Wa b ed oar ge d or... M or...
Getting Started Wand Emulation Connection — Connect the scanner to the appropriated port on the terminal/PC using the correct cable for the interface type you ordered. Reference Figure 2. IBM Connection — Connect the scanner to the IBM port on the ter- minal/PC using the correct IBM cable. Reference Figure 2. Keyboard Wedge Connection — The Keyboard Wedge cable has a ‘Y’ connection from the scanner.
Chapter 2 General Features Double Read Timeout The Double Read Timeout feature sets a time limit that determines how much time must pass before reading the same label again (e.g. two identical items in succession). START DURATION BAR CODE 0.1 Second 0.2 Second 0.3 Second 0.
General Features Double Read Timeout — continued START DURATION BAR CODE 0.5 Second 0.6 Second 0.7 Second 0.8 Second 0.
Powerdown Timeout Powerdown Timeout The Powerdown Timeout feature sets the time for automatically switching the unit off when the scanner is not in use.
General Features Powerdown Timeout — continued START DURATION BAR CODE 3 Minutes 4 Minutes END 2-4 PowerScan® 7000 SRI
LED and Beeper Indicators LED and Beeper Indicators Power On Alert Disables or enables the indication (from the Beeper) that the scanner is receiving power.
General Features LED Idle State Specifies state of scanner’s LED when the scanner is ready to read a bar code.
LED and Beeper Indicators Good Read: When to Indicate This feature specifies when the scanner will provide indication (beep and/or flash its green LED) upon successfully reading a bar code. Choices are: • Good Read = Indicate after decode • Good Read = Indicate after transmit • Good Read = Indicate after CTS goes inactive, then active This option, which uses CTS, is only valid for RS-232 interfaces.
General Features Good Read Beep Control This feature enables/disables the scanner’s ability to beep upon a successful decode of a bar code.
LED and Beeper Indicators Good Read Beep Frequency Adjusts the good read beep to sound at a selectable low, medium or high frequency, selectable from the list below. (Controls the beeper’s pitch/ tone.
General Features Good Read Beep Length Specifies the duration of a good read beep.
LED and Beeper Indicators Good Read Beep Length — continued START LENGTH BAR CODE 180msec 200msec END Product Reference Guide 2-11
General Features Good Read Beep Volume Selects the beeper volume (loudness) upon a good read beep. There are three selectable volume levels.
Scanning Features Scanning Features Scan Mode Selects the scan operating mode for the scanner.
General Features Scan Mode — continued START MODE BAR CODE Single Triggerless Stand Stand w/Illum END 2-14 PowerScan® 7000 SRI
Scanning Features Active Scanning Time This setting determines the amount of time the scanner continues to scan in triggerless or stand mode (see Scan Mode) once scanning has been activated.
General Features Active Scanning Time — continued START DURATION BAR CODE 2 Minutes 3 Minutes 4 Minutes END 2-16 PowerScan® 7000 SRI
Scanning Features Laser Pointer Control The Laser Pointer is a value-added option which might not have been included when your scanner was ordered. NOTE When the trigger is pressed and Scan Mode is set to Single, the laser pointer will be activated for the time period configured by this feature. Immediately following this, the scanner will start scanning. Disabling this feature turns the pointer off. START DURATION BAR CODE Disable 0.1 Seconds 0.2 Seconds 0.3 Seconds 0.
General Features Laser Pointer Control — continued START DURATION BAR CODE 0.5 Seconds 0.6 Seconds 0.8 Seconds 1 Second 1.2 Seconds 1.
Scanning Features Laser Pointer Control — continued START DURATION BAR CODE 2 Seconds 2.
General Features NOTES 2-20 PowerScan® 7000 SRI
Chapter 3 Interface Related Features At the time of this writing, the Scanner supports the interfaces listed in Table 3-1. Select the desired interface type from the table, then reference the page number given for the customizable features section associated with each interface. See Table 3-2 for a description of each Keyboard Wedge interface type (A through Z as listed).
Interface Related Features Table 3-1.
Table 3-2.
Interface Related Features Interface Selection START INTERFACE BAR CODE RS-232 Standard RS-232 Wincor-Nixdorf IBM 4683 Port 5B IBM 4683 Port 9B IBM 4683 Port 17 IBM USB USB Keyboard 3-4 PowerScan® 7000 SRI
Interface Selection Interface Selection — cont.
Interface Related Features Interface Selection — cont.
Interface Selection Interface Selection — cont.
Interface Related Features Interface Selection — cont.
Interface Features Interface Features Global Interface Features START STATE BAR CODE Obey Host Commands Ignore Host Commands Host Transmission Buffers = 1 Host Transmission Buffers = 2 END Product Reference Guide 3-9
Interface Related Features RS-232 Interface Features START BAUD RATE BAR CODE 1200 Baud 2400 Baud 4800 Baud 9600 Baud 19200 Baud 38400 Baud 3-10 PowerScan® 7000 SRI
Interface Features RS-232 Interface Features — cont.
Interface Related Features RS-232 Interface Features — cont.
Interface Features RS-232 Interface Features — cont.
Interface Related Features RS-232 Interface Features — cont. Hardware Flow Control Disable Hardware Control — The scanner transmits to the host regardless of any activity on the CTS line. Enable CTS Flow Control — The CTS signal controls transmission of data to the host. Enable CTS Scan Control — The CTS line must be active for the scanner to read and transmit data.
Interface Features RS-232 Interface Features — cont.
Interface Related Features RS-232 Interface Features — cont.
Interface Features RS-232 Interface Features — cont.
Interface Related Features RS-232 Interface Features — cont. Host Echo When enabled, this feature passes all data through the scanner to the host as it comes in. This feature is used for applications where “daisy chaining” of RS-232 devices onto the same cable is necessary.
Interface Features RS-232 Interface Features — cont.
Interface Related Features RS-232 Interface Features — cont.
Interface Features RS-232 Interface Features — cont. START STATE BAR CODE Signal Voltage: Normal RS-232 Signal Voltage: TTL Disable RS-232 Invert Enable RS-232 Invert END Beep on ASCII BEL Enables/disables ability of scanner to beep (sound a good read tone) on receiving an ASCII BEL (07 hex). Beep on Not on File Select for the host to beep (or not) when a not-on-file (host command) condition is detected by the host.
Interface Related Features ACK NAK Options This enables/disables the ability of the scanner to support the RS-232 ACK/NAK protocol. When configured, the scanner and/or host sends an “ACK” when it receives data properly, and sends “NAK” when the data is in error.
Interface Features ACK NAK Options — cont.
Interface Related Features RS-232 Interface Features — cont. ACK Character START MODE BAR CODE Sets the ACK character from the set of ASCII characters or any decimal value from 000 to 255. Pad entries of less than three digits with zeros, as in “005”. To configure this feature, scan the “START” bar code above to place the unit in Programming Mode, then the “Set ACK Character,” followed by the digits from the Alphanumeric table in Appendix C, AlphaNumeric Pad representing your desired character.
Interface Features RS-232 Interface Features — cont. NAK Character START MODE BAR CODE Sets the NAK character from the set of ASCII characters or any decimal value from 000 to 255. Pad entries of less than three digits with zeros, as in “005”. To configure this feature, scan the “START” bar code above to place the unit in Programming Mode, then the “Set NAK Character,” followed by the digits from the Alphanumeric table in Appendix C, AlphaNumeric Pad representing your desired character.
Interface Related Features RS-232 Interface Features — cont. Retry on ACK NAK Timeout Enables/disables retry after the configurable ACK NAK Timeout Value (set in the following feature) has expired.
Interface Features RS-232 Interface Features — cont. ACK NAK Timeout Value START MODE BAR CODE This item specifies the time the scanner will wait for an ACK character from the host following a label transmission.
Interface Related Features RS-232 Interface Features — cont. ACK NAK Retry Count START MODE BAR CODE This feature sets the number of times for the scanner to retry a label transmission under a retry condition.
Interface Features RS-232 Interface Features — cont. ACK NAK Error Handling This item specifies the method the scanner will use to handle errors detected while waiting to receive the ACK character from the host. Errors include unrecognized host commands and communication errors such as parity or framing errors.
Interface Related Features RS-232 Interface Features — cont. Transmission Failure Indication Enables/disables bad-label indication upon transmission failure.
Interface Features IBM-USB Interface Features IBM-USB Device usage The IBM-USB protocol allows for the scanner to be identified as one of two different types of bar code scanners. Depending on what other scanners you may already have connected to a IBM-USB POS, you may need to change this setting to enable all devices to communicate.
Interface Related Features IBM IBM Transmit Labels in Code 39 Format This feature enables/disables scanner's ability to set a symbology identifier for a specified label to Code 39 before transmitting that label data to an IBM host. This applies to: Code 128, Codabar and Code 93 for IBM USB; Code 128, Codabar and Code 93 for IBM Port 5B; and Codabar and Code 93 for IBM Port 9B.
Interface Features Wand Emulation Supported Symbologies The Wand Emulation interface will transmit bar code data as a wand device would. This interface will transmit the following bar code symbologies: • UPC/EAN • UPC/EAN with addons • Code 39 • Full ASCII Code 39 • Interleaved 2 of 5 • Codabar • Code 128 Pharmacode 39 is transmitted as Code 39, all other bar code symbology types read by the scanner will be transmitted as Code 128.
Interface Related Features Wand Emulation — cont. Bar/Space Polarity Low/High — Black will be transmitted as a low voltage level (0 to +0.7V) and space as high level (+2.4 to +5.25V). High/Low — Black will be transmitted as a high voltage level (+2.4 to +5.25V) and space as low level (0 to +0.7V). Wand Idle State This feature specifies the level of the wand output signal when idle. TTL logic levels: High voltage level (+2.4 to +5.25V) Low voltage level (0 to +0.7V).
Interface Features Wand Emulation — cont.
Interface Related Features Wand Emulation — cont. Signal Speed The speed of the transmission can be set. This selects the width of the minimum narrow bar.
Interface Features Wand Emulation — cont. Transmit Leading Noise Enables/disables the ability of scanner to generate noise transitions before label transitions in signals transmitted to the host.
Interface Related Features Wand Emulation — cont. Transmit Trailing Noise Enables/disables ability of scanner to generate noise transitions after label transitions in signals transmitted to the host.
Interface Features Symbology Conversion Wand Emulation can convert all bar codes to a single symbology.
Interface Related Features Symbology Conversion — cont.
Interface Features Keyboard Wedge/USB Keyboard As a keyboard interface, the scanner supports most popular PCs and IBM terminals. The installation of the wedge is a fairly simple process that doesn’t require any changes of software or hardware. NOTE All of the options in this section apply to the Keyboard Wedge, however, only Keyboard Layout, Caps Lock State and Control Characters apply to USB Keyboard. Keyboard Layout The Keyboard Layout option supports many countries.
Interface Related Features Keyboard Wedge/USB Keyboard — cont.
Interface Features Keyboard Wedge/USB Keyboard — cont.
Interface Related Features Keyboard Wedge/USB Keyboard — cont.
Interface Features Keyboard Wedge/USB Keyboard — cont. Caps Lock State Specifies which format the scanner sends character data.
Interface Related Features Keyboard Wedge/USB Keyboard — cont. Keyboard Simulation This feature does not apply to the USB Keyboard interface. NOTE All PCs check the keyboard status during the power-on Selftest. It is recommended that you enable this function if you are working without a keyboard installation. It simulates keyboard timing and passes the keyboard status to the PC during power-on. Control Characters Specifies how the scanner transmits ASCII control characters to the host.
Interface Features Keyboard Wedge/USB Keyboard — cont.
Interface Related Features Keyboard Wedge/USB Keyboard — cont. Wedge Quiet Interval This feature does not apply to the USB Keyboard interface. NOTE Quiet Interval is the amount of time to look for keyboard activity before the scanner breaks the keyboard connection in order to transmit data to the host.. START MODE BAR CODE Selectable from 000 (no interval) to 255 in 10 msec increments.
Interface Features Keyboard Wedge/USB Keyboard — cont. Intercharacter Delay START MODE BAR CODE One-half of the delay specified below is inserted between scancodes within each character. If the transmission speed is too high, the system may not be able to receive all characters. You may need to adjust the delay to make the system work properly. Selectable delay from 00 to 99 in 10msec increments.
Interface Related Features NOTES 3-50 PowerScan® 7000 SRI
Chapter 4 Data Editing Data Editing Overview It is not recommended to use these features with IBM or Wand Emulation interfaces. CAUTION When a bar code is scanned, additional information can be sent to the host computer along with the bar code data. This combination of bar code data and supplementary user-defined data is called a “message string.” The features in this chapter can be used to build specific user-defined data into a message string.
Data Editing Please Keep In Mind... • Modifying a message string is not a mandatory requirement. Data editing is sophisticated feature allowing highly customizable output for advanced users. Factory default settings for data editing is typically set to NONE. • A prefix or suffix may be applied (reference the Symbologies chapter for these settings) across all symbologies (set via the Global features in this chapter).
Global Prefix/Suffix Global Prefix/Suffix — continued Example: Setting a Prefix In this example, we’ll set a prefix for all symbologies. 1. Determine which ASCII character(s) are to be added to scanned bar code data. In this example, we’ll add a dollar sign (‘$’) as a prefix. 2. Scan the START bar code. 3. Scan the SET PREFIX bar code. 4. Reference the ASCII Chart on the inside back cover of this manual, to find the hex value assigned to the desired character.
Data Editing Global Prefix/Suffix — continued START MODE BAR CODE Sets up to 20 characters each from the set of ASCII characters or any hex value from 0 to FF. To configure this feature, scan the “START” bar code above to place the unit in Programming Mode, then the “Set Prefix” or “Set Suffix,” followed by the digits (in hex) from the Alphanumeric table in Appendix C, AlphaNumeric Pad representing your desired character(s). Reference the section, Example: Setting a Prefix, for more information.
AIM ID AIM ID AIM label identifiers (as opposed to custom characters you select yourself as with label identifiers) can be included with scanned bar code data. AIM label identifiers consist of three characters as follows: • A close brace character (ASCII ‘]’), followed by...
Data Editing AIM ID — continued START STATE BAR CODE Disable AIM ID Enable AIM ID END 4-6 PowerScan® 7000 SRI
Label ID Label ID A Label ID is a customizable code of up to two ASCII characters (00FF), used to identify a bar code (symbology) type. It can be appended previous to or following the transmitted bar code data depending upon how this option is enabled. This feature provides options for configuring custom Label IDs individually per symbology. If you wish to program the scanner to always include an industry standard label identifier for ALL symbology types, see the previous feature, AIM ID.
Data Editing Label ID — continued START OPTION BAR CODE Label ID Transmission: Disable Label ID Position: Before Bar Code Data Label ID Position: After Bar Code Data Set UPC-A Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: A (41 hex) Set UPC-A w/P2 Addon Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: A (41 hex) Set UPC-A w/P5 Addon Label ID Character(s) 4-8 DEFAULT SETTING FOR THIS FEATURE: A (41 hex) PowerScan® 7000 SRI
Label ID Label ID — continued START OPTION BAR CODE Set UPC-A w/C128 Addon Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: A (41 hex) Set UPC-E Label ID Character(s) Set UPC-E w/P2 Addon Label ID Character(s) Set UPC-E w/P5 Addon Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: E (45 hex) DEFAULT SETTING FOR THIS FEATURE: E (45 hex) DEFAULT SETTING FOR THIS FEATURE: E (45 hex) Set UPC-E w/C128 Addon Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: E (45 hex) Set EAN-8 Label ID C
Data Editing Label ID — continued START OPTION Set EAN-8 w/P5 Addon Label ID Character(s) Set EAN-8 w/C128 Addon Label ID Character(s) Set EAN-13 Label ID Character(s) BAR CODE DEFAULT SETTING FOR THIS FEATURE: FF (4646 hex) DEFAULT SETTING FOR THIS FEATURE: FF (4646 hex) DEFAULT SETTING FOR THIS FEATURE: F (46 hex) Set EAN-13 w/P2 Addon Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: F (46 hex) Set EAN-13 w/P5 Addon Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: F (46 hex) Set EAN-1
Label ID Label ID — continued START OPTION Set GTIN Label ID Character(s) Set GTIN w/P2 addon Label ID Character(s) Set GTIN w/P5 addon Label ID Character(s) Set GTIN w/C128 addon Label ID Character(s) Set RSS-14 Label ID Character(s) Set RSS-14 Composite Label ID Character(s) BAR CODE DEFAULT SETTING FOR THIS FEATURE: G (47 hex) DEFAULT SETTING FOR THIS FEATURE: G2 (4732 hex) DEFAULT SETTING FOR THIS FEATURE: G5 (4735 hex) DEFAULT SETTING FOR THIS FEATURE: G8 (4738 hex) DEFAULT SETTING FOR THIS
Data Editing Label ID — continued START OPTION Set RSS Expanded Label ID Character(s) Set RSS Expanded Composite Label ID Character(s) Set RSS Limited Label ID Character(s) Set RSS Limited Composite Label ID Character(s) Set Code 39 Label ID Character(s) 4-12 BAR CODE DEFAULT SETTING FOR THIS FEATURE: RX (5258 hex) DEFAULT SETTING FOR THIS FEATURE: 5258 Hex (RX) DEFAULT SETTING FOR THIS FEATURE: 524C Hex (RL) DEFAULT SETTING FOR THIS FEATURE: 524C Hex (RL) DEFAULT SETTING FOR THIS FEATURE: * (2A
Label ID Label ID — continued START OPTION BAR CODE Set Pharmacode 39 Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: A (41 hex) Set Code 128 Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: # (23 hex) Set I 2 of 5 Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: i (69 hex) Set Codabar Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: % (25 hex) Set Code 93 Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: & (26 hex) Set Code 11 Label ID Character(s) DEFAULT SETTING
Data Editing Label ID — continued START OPTION Set MSI/Plessey Label ID Character(s) Set Std 2 of 5 Label ID Character(s) Set PDF 417 Label ID Character(s) BAR CODE DEFAULT SETTING FOR THIS FEATURE: @ (40 hex) DEFAULT SETTING FOR THIS FEATURE: i (69 hex) DEFAULT SETTING FOR THIS FEATURE: P (50 hex) Set Micro PDF 417 Label ID Character(s) DEFAULT SETTING FOR THIS FEATURE: 6D50 Hex (mP) END 4-14 PowerScan® 7000 SRI
Label ID Case Conversion This feature can convert scanned bar code data to either all lower case or all upper case characters. Case conversion affects ONLY scanned bar code data, and does not affect Label ID, Prefix, Suffix, or other appended data.
Data Editing Character Conversion Character conversion is an eight byte configuration item. The eight bytes are 4 character pairs represented in hexadecimal ASCII values. The first character in the pair is the character that will be converted. The second character in the pair is the character to convert to. If the character to convert in a pair is FF, then no conversion is done.
Label ID Character Conversion — continued START MODE Character Conversion BAR CODE DEFAULT SETTING FOR THIS FEATURE: FFFFFFFFFFFFFFFF hex (no conversion) END Product Reference Guide 4-17
Data Editing NOTES 4-18 PowerScan® 7000 SRI
Chapter 5 Symbologies The scanner supports the following symbologies (bar code types). Options for each symbology are included in this chapter. ·UPC-A ·Pharmacode 39 ·UPC-E ·Code 128 and EAN 128 ·EAN-13 ·Interleaved 2 of 5 ·EAN-8 ·Codabar ·RSS-14 ·Code 93 ·RSS Expanded ·Code 11 ·RSS Limited ·MSI/Plessey ·Code 39 ·Standard 2 of 5 2D Symbologies ·PDF 417 ·Micro PDF 417 Factory Defaults — are indicated in bold text and are highlighted with yellow throughout this section.
Symbologies UPC-A The following options apply to the UPC-A symbology. Disable/Enable UPC-A When disabled, the scanner will not read UPC-A bar codes. Check Digit Transmission Enable this option to transmit the check digit along with UPC-A bar code data. Expand UPC-A to EAN-13 Expands UPC-A data to the EAN-13 data format. Selecting this feature also changes the symbology ID to match those required for EAN-13. System Number Transmission This feature enables/disables transmission of UPC-A System Number.
UPC-A UPC-A — continued START STATE BAR CODE Disable UPC-A Enable UPC-A Don’t Send Check Digit Send Check Digit Don’t Expand to EAN-13 Expand to EAN-13 Product Reference Guide 5-3
Symbologies UPC-A — continued START STATE BAR CODE Disable System Number Transmission Enable System Number Transmission END 5-4 PowerScan® 7000 SRI
UPC-A UPC-A — continued UPC-A Minimum Reads This feature specifies the minimum number of consecutive times a UPCA label must be decoded before it is accepted as good read.
Symbologies UPC-A — continued In-store Minimum Reads This feature specifies the minimum number of consecutive times an instore printed label must be decoded before it is accepted as good read.
UPC-A UPC-A — continued Add-On Timer This feature sets the period of time after a UPC-A label is decoded when optional add-on reading is enabled. START STATE BAR CODE This item is used to set the timer value when optional add-ons are enabled. The selectable range is from 010 to 300 mSec in 10 mSec increments.
Symbologies UPC-E The following options apply to the UPC-E symbology. Disable/Enable UPC-E When disabled, the scanner will not read UPC-E bar codes. Check Digit Transmission Enable this option to transmit the check digit along with UPC-E bar code data. System Number The System Number (SN) which is usually a zero (0) in the leading position can be optionally included (or not) with scanned bar code data. Expand UPC-E to UPC-A Enables/disables expansion of UPC-E labels to UPC-A.
UPC-E UPC-E — continued START STATE BAR CODE Disable UPC-E Enable UPC-E Don’t Send Check Digit Send Check Digit Exclude System Number Include System Number Product Reference Guide 5-9
Symbologies UPC-E — continued START STATE BAR CODE Don’t Expand UPC-E to UPC-A Expand UPC-E to UPC-A Don’t Expand UPC-E to EAN-13 Expand UPC-E to EAN-13 END 5-10 PowerScan® 7000 SRI
UPC-E UPC-E — continued Minimum Reads This feature specifies the minimum number of consecutive times a UPCE label must be decoded before it is accepted as good read.
Symbologies GTIN The following options apply to the GTIN label data format. Disable/Enable GTIN When this feature is enabled, the scanner will translate UPC/EAN labels to the 14 digit GTIN format.
EAN-13 EAN-13 The following options apply to the EAN-13 symbology. Disable/Enable EAN-13 When disabled, the scanner will not read EAN-13 bar codes. Check Digit Transmission Enable this option to transmit the check digit along with EAN-13 bar code data. EAN-13 Flag 1 Character Enables/disables transmission of an EAN/JAN13 Flag1 character. ISBN When enabled, this feature truncates the leading three digits from labels that contain ISBN (International Standard Book Number).
Symbologies EAN-13 — continued START STATE BAR CODE Disable EAN-13 Enable EAN-13 Don’t Send Check Digit Send Check Digit Don’t Transmit EAN-13 Flag 1 Char Transmit EAN-13 Flag 1 Char 5-14 PowerScan® 7000 SRI
EAN-13 EAN-13 — continued START STATE BAR CODE Disable ISBN Enable ISBN END Product Reference Guide 5-15
Symbologies EAN-13 — continued Minimum Reads This feature specifies the minimum number of consecutive times an EAN-13 label must be decoded before it is accepted as good read.
EAN-8 EAN-8 The following options apply to the EAN-8 symbology. Disable/Enable EAN-8 When disabled, the scanner will not read EAN-8 bar codes. Check Digit Transmission Enable this option to transmit the check Digit along with EAN-8 bar code data.
Symbologies EAN-8 — continued Expand EAN-8 to EAN-13 — Expands EAN-8 data to the EAN-13 data format. Selecting this feature also changes the symbology ID to match those required for EAN-13.
EAN-8 EAN-8 — continued Minimum Reads This feature specifies the minimum number of consecutive times an EAN-8 label must be decoded before it is accepted as good read.
Symbologies Add-ons Add-ons (or supplemental characters) are commonly added to the end of UPC/EAN bar codes. The scanner will read the add-ons if they are enabled and in the field of view. Three add-on types are supported: 2digit, 5-digit and Code 128 add-ons. Supported options are: None — This option directs the scanner to ignore add-on portion of a UPC/EAN bar code but still read the main portion of the bar code. 2 Digits — The scanner will optionally read 2-digit add-ons with the UPC/EAN label.
Add-ons Add-ons — continued START STATE BAR CODE Disable Optional 2Digit Add-ons Enable Optional 2-Digit Add-ons Disable Optional 5Digit Add-ons Enable Optional 5-Digit Add-ons Disable Optional Code 128 Add-ons Enable Optional Code 128 Add-ons END Product Reference Guide 5-21
Symbologies RSS-14 The following options apply to the RSS-14 symbology. Disable/Enable RSS-14 When this feature is disabled, the scanner will not read RSS-14 bar codes. UCC/EAN 128 Emulation When enabled, RSS-14 bar codes will be translated to the UCC/EAN 128 label data format.
RSS-14 RSS-14 — continued START STATE BAR CODE Disable RSS-14 Enable RSS-14 Disable UCC/EAN 128 Emulation Enable UCC/EAN 128 Emulation END Product Reference Guide 5-23
Symbologies RSS-14 — continued Minimum Reads This feature specifies the minimum number of consecutive times an RSS-14 label must be decoded before it is accepted as good read.
RSS-14 RSS-14 — continued RSS-14 2D Component This feature controls the requirement that a 2D label component be decoded when an RSS-14 base label is decoded. • Enable — Requires that a 2D label component be decoded when an RSS-14 base label is decoded. • Disable — Does NOT require a 2D label component to be decoded when an RSS-14 base label is decoded.
Symbologies RSS Expanded The following options apply to the RSS Expanded symbology. Disable/Enable RSS Expanded When this feature is disabled, the scanner will not read RSS Expanded bar codes. UCC/EAN 128 Emulation When enabled, RSS Expanded bar codes will be translated to the UCC/ EAN 128 label data format.
RSS Expanded RSS Expanded — continued Length Control Fixed Length Decoding — When fixed length decoding is enabled, the scanner will decode a bar code if the label length matches one of the configurable fixed lengths. Variable Length Decoding — When variable length decoding is enabled, the scanner will decode a bar code if the label length falls in the range of the configurable minimum and maximum length. Configuring Fixed Length Decoding: 1. Scan the START bar code. 2.
Symbologies RSS Expanded — continued START MODE BAR CODE Variable Length Decoding Fixed Length Decoding END 5-28 PowerScan® 7000 SRI
RSS Expanded RSS Expanded — continued RSS Expanded Length 1, Length 2 Programming Instructions 1. Scan the START bar code. 2. Scan either the Set Length 1 or Set Length 2 bar code. 3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representing the length in decimal notation. For RSS Expanded bar codes, only the data characters are included in the length calculations. NOTE 4. Scan the END bar code.
Symbologies RSS Expanded — continued Minimum Reads This feature specifies the minimum number of consecutive times an RSS Expanded label must be decoded before it is accepted as good read.
RSS Expanded RSS Expanded — continued RSS Expanded 2D Component This feature controls the requirement that a 2D label component be decoded when an RSS Expanded base label is decoded. • Enable — Requires that a 2D label component be decoded when an RSS Expanded base label is decoded. • Disable — Does NOT require a 2D label component to be decoded when an RSS Expanded base label is decoded.
Symbologies RSS Limited The following options apply to the RSS Limited symbology. Disable/Enable RSS Limited When this feature is disabled, the scanner will not read RSS Limited bar codes. UCC/EAN 128 Emulation When enabled, RSS Limited bar codes will be translated to the UCC/ EAN 128 label data format.
RSS Limited RSS Limited — continued Minimum Reads This feature specifies the minimum number of consecutive times an RSS Limited label must be decoded before it is accepted as good read.
Symbologies RSS Limited — continued RSS Limited 2D Component This feature controls the requirement that a 2D label component be decoded when an RSS Limited base label is decoded. • Enable — Requires that a 2D label component be decoded when an RSS Limited base label is decoded. • Disable — Does NOT require a 2D label component to be decoded when an RSS Limited base label is decoded.
Code 39 Code 39 The following options apply to the Code 39 symbology. Disable/Enable Code 39 When this feature is disabled, the scanner will not read Code 39 bar codes. Check Character Calculation When enabled, the scanner will calculate the check character of the labels. Turn this option on only when a checksum is present in the Code 39 labels. Check Character Transmit Enable this option to transmit the check character with scanned bar code data.
Symbologies Code 39 — continued START STATE BAR CODE Disable Code 39 Enable Code 39 Disable Check Char Calculation Enable Check Char Calculation Disable Check Char Transmission Enable Check Char Transmission 5-36 PowerScan® 7000 SRI
Code 39 Code 39 — continued START STATE BAR CODE Don’t Transmit Start/Stop Characters Transmit Start/Stop Characters Disable Code 39 Full ASCII Enable Code 39 Full ASCII END Product Reference Guide 5-37
Symbologies Code 39 — continued Length Control Fixed Length Decoding — When fixed length decoding is enabled, the scanner will decode a bar code if the label length matches one of the configurable fixed lengths. Variable Length Decoding — When variable length decoding is enabled, the scanner will decode a bar code if the label length falls in the range of the configurable minimum and maximum length. Configuring Fixed Length Decoding: 1. Scan the START bar code. 2. Scan the Fixed Length Decoding bar code.
Code 39 Code 39 — continued START MODE BAR CODE Variable Length Decoding Fixed Length Decoding END Product Reference Guide 5-39
Symbologies Code 39 Length 1, Length 2 Programming Instructions 1. Scan the START bar code. 2. Scan either the Set Length 1 or Set Length 2 bar code. 3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representing the length in decimal notation. For Code 39 bar codes, all check, data and full ASCII shift characters are included in the length calculations. Start/Stop characters are not included. NOTE 4. Scan the END bar code.
Code 39 Code 39 — continued Quiet Zones This feature enables/disables the requirement that quiet zones must be present for Code 39 bar codes.
Symbologies Code 39 — continued Code 39 Stitching Enables/disables stitching for Code 39 labels. When parts of a Code 39 bar code are presented to the scanner with this feature enabled, the bar code parts will be assembled by the scanner’s software, and the data will be decoded if all bar code proofing requirements are met.
Code 39 Code 39 — continued Minimum Reads This feature specifies the minimum number of consecutive times an Code 39 label must be decoded before it is accepted as good read.
Symbologies Pharmacode 39 The following options apply to the Pharmacode 39 symbology. Disable/Enable Pharmacode 39 When this feature is disabled, the scanner will not read Pharmacode 39 bar codes.
Pharmacode 39 Pharmacode 39 — continued Start/Stop Characters Enables or disables transmission of Pharmacode 39 start/stop characters. Check Character Transmit Enable this option to transmit the check character with scanned bar code data.
Symbologies Code 128 and EAN 128 The following options apply to the Code 128 and EAN 128 symbologies. Code 128 — When this feature is disabled, the scanner will not read Code 128 bar codes. UCC/EAN 128 — Enables/disables ability of scanner to translate UCC/ EAN 128 labels to the Code 128 data format.
Code 128 and EAN 128 Code 128 and EAN 128 — continued Transmit Function Characters Enables/disables transmission of Code128 function characters 1, 2, 3, and 4.
Symbologies Code 128 and EAN 128 — continued Length Control Fixed Length Decoding — When fixed length decoding is enabled, the scanner will decode a bar code if the label length matches one of the configurable fixed lengths. Variable Length Decoding — When variable length decoding is enabled, the scanner will decode a bar code if the label length falls in the range of the configurable minimum and maximum length. Configuring Fixed Length Decoding: 1. Scan the START bar code. 2.
Code 128 and EAN 128 Code 128 and EAN 128 — continued START MODE BAR CODE Variable Length Decoding Fixed Length Decoding END Product Reference Guide 5-49
Symbologies Code 128 and EAN 128 — continued Code 128 Length 1, Length 2 Programming Instructions 1. Scan the START bar code. 2. Scan either the Set Length 1 or Set Length 2 bar code. 3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representing the length in decimal notation. For Code 128 bar codes, only the data characters are included in the length calculations. NOTE 4. Scan the END bar code.
Code 128 and EAN 128 Code 128 and EAN 128 — continued Code 128 Stitching Enables/disables stitching for Code 128 labels. When parts of a Code 128 bar code are presented to the scanner with this feature enabled, the bar code parts will be assembled by the scanner’s software, and the data will be decoded if all bar code proofing requirements are met.
Symbologies Code 128 and EAN 128 — continued Minimum Reads This feature specifies the minimum number of consecutive times an Code 128 label must be decoded before it is accepted as good read.
Interleaved 2 of 5 Interleaved 2 of 5 The following options apply to the Interleaved 2 of 5 (I 2 of 5) symbology. Disable/Enable Interleaved 2 of 5 When this feature is disabled, the scanner will not read Interleaved 2 of 5 bar codes. Check Digit Calculation When enabled, the scanner will calculate the check digit of the labels. Check Digit Transmit Enable this option to transmit the check digit with scanned bar code data.
Symbologies Interleaved 2 of 5 — continued START STATE BAR CODE Disable Interleaved 2 of 5 Enable Interleaved 2 of 5 Disable Check Digit Calculation Enable Check Digit Calculation Disable Check Digit Transmission Enable Check Digit Transmission END 5-54 PowerScan® 7000 SRI
Interleaved 2 of 5 Interleaved 2 of 5 — continued Length Control Fixed Length Decoding — When fixed length decoding is enabled, the scanner will decode a bar code if the label length matches one of the configurable fixed lengths. Variable Length Decoding — When variable length decoding is enabled, the scanner will decode a bar code if the label length falls in the range of the configurable minimum and maximum length. Configuring Fixed Length Decoding: 1. Scan the START bar code. 2.
Symbologies Interleaved 2 of 5 — continued START MODE BAR CODE Variable Length Decoding Fixed Length Decoding END 5-56 PowerScan® 7000 SRI
Interleaved 2 of 5 Interleaved 2 of 5 — continued Interleaved 2 of 5 Length 1, Length 2 Programming Instructions 1. Scan the START bar code. 2. Scan either the Set Length 1 or Set Length 2 bar code. 3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits representing the length in decimal notation. NOTE For Interleaved 2 of 5 bar codes, lengths must be an even number. Additionally, all check and data characters are included in the length calculations. 4. Scan the END bar code.
Symbologies Interleaved 2 of 5 — continued Interleaved 2 of 5 Stitching Enables/disables stitching for Interleaved 2 of 5 labels. When parts of an Interleaved 2 of 5 bar code are presented to the scanner with this feature enabled, the bar code parts will be assembled by the scanner’s software, and the data will be decoded if all bar code proofing requirements are met.
Interleaved 2 of 5 Interleaved 2 of 5 — continued Minimum Reads This feature specifies the minimum number of consecutive times an Interleaved 2 of 5 label must be decoded before it is accepted as good read.
Symbologies Codabar The following options apply to the Codabar symbology. Disable/Enable Codabar When this feature is disabled, the scanner will not read Codabar bar codes. Check Character Verification When enabled, the scanner will verify the check character of the labels. Check Character Transmit Enable this option to transmit the check character with scanned bar code data.
Codabar Codabar — continued START STATE BAR CODE Disable Codabar Enable Codabar Disable Check Char Verification Enable Check Char Verification Disable Check Char Transmission Enable Check Char Transmission END Product Reference Guide 5-61
Symbologies Codabar — continued Length Control Fixed Length Decoding — When fixed length decoding is enabled, the scanner will decode a bar code if the label length matches one of the configurable fixed lengths. Variable Length Decoding — When variable length decoding is enabled, the scanner will decode a bar code if the label length falls in the range of the configurable minimum and maximum length. Configuring Fixed Length Decoding: 1. Scan the START bar code. 2. Scan the Fixed Length Decoding bar code.
Codabar Codabar — continued START MODE BAR CODE Variable Length Decoding Fixed Length Decoding END Product Reference Guide 5-63
Symbologies Codabar — continued Codabar Length 1, Length 2 Programming Instructions 1. Scan the START bar code. 2. Scan either the Set Length 1 or Set Length 2 bar code. 3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representing the length in decimal notation. For Codabar bar codes, all start, stop, check and data characters are included in the length calculations. NOTE 4. Scan the END bar code.
Codabar Codabar — continued Quiet Zones This feature enable/disables the requirement that quiet zones must be present for Codabar bar codes.
Symbologies Codabar — continued Start/Stop Character Type Codabar has four pairs of Start/Stop patterns. Select one pair to match your application. Start/Stop Character Transmission The transmission of start and end characters of Codabar is selected below. Start/Stop Character Match This feature enables/disables the requirement that start and stop characters match.
Codabar Codabar — continued START STATE BAR CODE Start/Stop Type: ABCD/ TN*E Start/Stop Type: ABCD/ ABCD Start/Stop Type: abcd/tn*e Start/Stop Type: abcd/ abcd Disable Start/Stop Char Transmission Enable Start/Stop Char Transmission Product Reference Guide 5-67
Symbologies Codabar — continued START STATE BAR CODE Disable Start/Stop Char Match Enable Start/Stop Char Match END 5-68 PowerScan® 7000 SRI
Codabar Codabar — continued Codabar Stitching Enables/disables stitching for Codabar labels. When parts of a Codabar label are presented to the scanner with this feature enabled, the bar code parts will be assembled by the scanner’s software, and the data will be decoded if all bar code proofing requirements are met.
Symbologies Codabar — continued Minimum Reads This feature specifies the minimum number of consecutive times an Codabar label must be decoded before it is accepted as good read.
Code 93 Code 93 The following options apply to the Code 93 symbology. Disable/Enable Code 93 When this feature is disabled, the scanner will not read Code 93 bar codes.
Symbologies Code 93 — continued Length Control Fixed Length Decoding — When fixed length decoding is enabled, the scanner will decode a bar code if the label length matches one of the configurable fixed lengths. Variable Length Decoding — When variable length decoding is enabled, the scanner will decode a bar code if the label length falls in the range of the configurable minimum and maximum length. Configuring Fixed Length Decoding: 1. Scan the START bar code. 2. Scan the Fixed Length Decoding bar code.
Code 93 Code 93 — continued START MODE BAR CODE Variable Length Decoding Fixed Length Decoding END Product Reference Guide 5-73
Symbologies Code 93 — continued Code 93 Length 1, Length 2 Programming Instructions 1. Scan the START bar code. 2. Scan either the Set Length 1 or Set Length 2 bar code. 3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representing the length in decimal notation. For Code 93 bar codes, only the data characters are included in the length calculations. NOTE 4. Scan the END bar code.
Code 93 Code 93 — continued Code 93 Stitching Enables/disables stitching for Code 93 bar codes. When parts of a Code 93 label are presented to the scanner with this feature enabled, the bar code parts will be assembled by the scanner’s software, and the data will be decoded if all bar code proofing requirements are met.
Symbologies Code 93 — continued Minimum Reads This feature specifies the minimum number of consecutive times an Code 93 label must be decoded before it is accepted as good read.
Code 11 Code 11 The following options apply to the Code 11 symbology. Disable/Enable Code 11 When this feature is disabled, the scanner will not read Code 11 bar codes.
Symbologies Code 11 — continued Number of Check Characters Specifies the number of Code 11 check characters (one or two) to be calculated and verified. Check Character Transmit Enables/disables transmission of Code 11 check characters.
Code 11 Code 11 — continued Length Control Fixed Length Decoding — When fixed length decoding is enabled, the scanner will decode a bar code if the label length matches one of the configurable fixed lengths. Variable Length Decoding — When variable length decoding is enabled, the scanner will decode a bar code if the label length falls in the range of the configurable minimum and maximum length. Configuring Fixed Length Decoding: 1. Scan the START/END bar code. 2. Scan the Fixed Length Decoding bar code.
Symbologies Code 11 — continued START/END MODE BAR CODE Variable Length Decoding Fixed Length Decoding 5-80 PowerScan® 7000 SRI
Code 11 Code 11 — continued Code 11 Length 1, Length 2 Programming Instructions 1. Scan the START/END bar code. 2. Scan either the Set Length 1 or Set Length 2 bar code. 3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representing the length in decimal notation. For Code 11 bar codes, only the data characters are included in the length calculations. NOTE 4. Scan the START/END bar code.
Symbologies Code 11 — continued Minimum Reads This feature specifies the minimum number of consecutive times an Code 11 label must be decoded before it is accepted as good read.
MSI/Plessey MSI/Plessey The following options apply to the MSI/Plessey symbology. Disable/Enable MSI/Plessey When this feature is disabled, the scanner will not read MSI/Plessey bar codes. Check Digit Verification This feature specifies whether one or two check digits are to be calculated and verified. Check Digit Transmit When this option is enabled, the scanner will transmit one-digit or twodigit check digits, depending upon the setting for check digit verification.
Symbologies MSI/Plessey — continued START STATE BAR CODE Disable MSI/Plessey Enable MSI/Plessey Disable Check Digit Verification Enable Check Digit Verification 1-Digit Check Digit Verification 2-Digit Check Digit Verification 5-84 PowerScan® 7000 SRI
MSI/Plessey MSI/Plessey — continued START STATE BAR CODE Disable Check Digit Transmission Enable Check Digit Transmission END Product Reference Guide 5-85
Symbologies MSI/Plessey — continued Length Control Fixed Length Decoding — When fixed length decoding is enabled, the scanner will decode a bar code if the label length matches one of the configurable fixed lengths. Variable Length Decoding — When variable length decoding is enabled, the scanner will decode a bar code if the label length falls in the range of the configurable minimum and maximum length. Configuring Fixed Length Decoding: 1. Scan the START bar code. 2.
MSI/Plessey MSI/Plessey — continued START MODE BAR CODE Variable Length Decoding Fixed Length Decoding END Product Reference Guide 5-87
Symbologies MSI/Plessey — continued MSI/Plessey Length 1, Length 2 Programming Instructions 1. Scan the START bar code. 2. Scan either the Set Length 1 or Set Length 2 bar code. 3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representing the length in decimal notation. For MSI/Plessey bar codes, all check and data characters are included in the length calculations. NOTE Scan the END bar code.
MSI/Plessey MSI/Plessey — continued MSI/Plessey Stitching Enables/disables stitching for MSI/Plessey bar codes. When parts of an MSI/Plessey label are presented to the scanner with this feature enabled, the bar code parts will be assembled by the scanner’s software, and the data will be decoded if all bar code proofing requirements are met.
Symbologies MSI/Plessey — continued Minimum Reads This feature specifies the minimum number of consecutive times an MSI/Plessey label must be decoded before it is accepted as good read.
MSI/Plessey Standard 2 of 5 The following options apply to the Standard 2 of 5 symbology. Disable/Enable Standard 2 of 5 When this feature is disabled, the scanner will not read Standard 2 of 5 bar codes. Check Digit Verification When enabled, the scanner will verify the check digit of the labels. Check Digit Transmit When this option is enabled, the scanner will transmit the check digit.
Symbologies Standard 2 of 5 — continued START STATE BAR CODE Disable Std 2 of 5 Enable Std 2 of 5 Disable Check Digit Verification Enable Check Digit Verification Disable Check Digit Transmission Enable Check Digit Transmission END 5-92 PowerScan® 7000 SRI
MSI/Plessey Standard 2 of 5 — continued Length Control Fixed Length Decoding — When fixed length decoding is enabled, the scanner will decode a bar code if the label length matches one of the configurable fixed lengths. Variable Length Decoding — When variable length decoding is enabled, the scanner will decode a bar code if the label length falls in the range of the configurable minimum and maximum length. Configuring Fixed Length Decoding: 1. Scan the START bar code. 2.
Symbologies Standard 2 of 5 — continued START MODE BAR CODE Variable Length Decoding Fixed Length Decoding END 5-94 PowerScan® 7000 SRI
MSI/Plessey Standard 2 of 5 — continued Standard 2 of 5 Length 1, Length 2 Programming Instructions 1. Scan the START bar code. 2. Scan either the Set Length 1 or Set Length 2 bar code. 3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representing the length in decimal notation. For Standard 2 of 5 bar codes, all check and data characters are included in the length calculations. NOTE 4. Scan the END bar code.
Symbologies Standard 2 of 5 — continued Standard 2 of 5 Stitching Enables/disables stitching for Standard 2 of 5 bar codes. When parts of a Standard 2 of 5 label are presented to the scanner with this feature enabled, the bar code parts will be assembled by the scanner’s software, and the data will be decoded if all bar code proofing requirements are met.
MSI/Plessey Standard 2 of 5 — continued Minimum Reads This feature specifies the minimum number of consecutive times a Standard 2 of 5 label must be decoded before it is accepted as good read.
Symbologies PDF 417 The following options apply to the PDF 417 symbology. Disable/Enable PDF 417 When this feature is disabled, the scanner will not read PDF 417 bar codes.
PDF 417 PDF 417 — continued Length Control Fixed Length Decoding — When fixed length decoding is enabled, the scanner will decode a bar code if the label length matches one of the configurable fixed lengths. Variable Length Decoding — When variable length decoding is enabled, the scanner will decode a bar code if the label length falls in the range of the configurable minimum and maximum length. Configuring Fixed Length Decoding: 1. Scan the START bar code. 2. Scan the Fixed Length Decoding bar code. 3.
Symbologies PDF 417 — continued START MODE BAR CODE Variable Length Decoding Fixed Length Decoding END 5-100 PowerScan® 7000 SRI
PDF 417 PDF 417 — continued PDF 417 Length 1, Length 2 Programming Instructions 1. Scan the START bar code. 2. Scan either the Set Length 1 or Set Length 2 bar code. 3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representing the length in decimal notation. For PDF 417 bar codes, only the data characters are included in the length calculations. NOTE Scan the END bar code.
Symbologies PDF 417 — continued Minimum Reads This feature specifies the minimum number of consecutive times a PDF 417 label must be decoded before it is accepted as good read.
Micro PDF 417 Micro PDF 417 The following options apply to the Micro PDF 417 symbology. Disable/Enable Micro PDF 417 When this feature is disabled, the scanner will not read Micro PDF 417 bar codes.
Symbologies Micro PDF 417 — continued PDF 128 Emulation Depending upon the other configuration features listed below, this feature specifies which AIM ID to use for Micro PDF 417 labels when performing EAN 128 or Code 128 emulation. • • If this feature (PDF 128 Emulation) is enabled, and the feature AIM ID is enabled while the scanner is performing EAN 128 or Code 128 emulation for a Micro PDF 417 label, the appropriate EAN 128 or Code 128 AIM ID is appended to the label data.
Micro PDF 417 Micro PDF 417 — continued Length Control Fixed Length Decoding — When fixed length decoding is enabled, the scanner will decode a bar code if the label length matches one of the configurable fixed lengths. Variable Length Decoding — When variable length decoding is enabled, the scanner will decode a bar code if the label length falls in the range of the configurable minimum and maximum length. Configuring Fixed Length Decoding: 1. Scan the START/END bar code. 2.
Symbologies Micro PDF 417 — continued START/END MODE BAR CODE Variable Length Decoding Fixed Length Decoding 5-106 PowerScan® 7000 SRI
Micro PDF 417 Micro PDF 417 — continued Micro PDF 417 Length 1, Length 2 Programming Instructions 1. Scan the START/END bar code. 2. Scan either the Set Length 1 or Set Length 2 bar code. 3. Turn to Appendix C, Alpha-Numeric Pad and scan the three digits (zero padded) representing the length in decimal notation. For Micro PDF 417 bar codes, only the data characters are included in the length calculations. NOTE Scan the START/END bar code.
Symbologies Micro PDF 417 — continued Minimum Reads This feature specifies the minimum number of consecutive times a Micro PDF 417 label must be decoded before it is accepted as good read.
Appendix A Product Specifications Optical and Read Performance Parameters Parameter Scanning Width Minimum Resolution Depth of Field (75% read rate; 90% PCS; Code 39 with 2.5:1 W/N ratio except for 13 mils) Specification 2” wide at 1” from scanner 6” wide at 7” from scanner 4mil minimum element size (at some distance, no implied DOF) 5 mil — 3 to 6”/7.6 to 15.2 cm 7.5 mil — 2 to 16”/5.1 to 40.6 cm 10 mil — 1.5 to 22”/3.8 to 55.9 cm 13 mil — 1 to 30”/2.5 to 76.2 cm 20 mil — 1 to 42”/2.5 to 106.
Physical Properties Parameter Specification Dimensions (Typical): Height 7.5”/190 mm Length 4.5”/115 mm Width 3.0”/75 mm Weight 10 ounces/284 g (without cable) Voltage Requirements +4 to 14 VDC +/- 5% Electrical Parameters Parameter Specification Input Voltage Input Power Maximum Operating Power Typical Operating Power Typical Standby Power Input Current Maximum Operating Current Typical Operating Current Typical Standby Current Sleep Mode Current A-2 +4 to 14 VDC 2.4 W 1.
Environmental Parameters Environmental Parameters Parameter Mechanical Shock Contaminants Spray/rain Dust/particulate Specification 50 drops from 6.
NOTES A-4 PowerScan® 7000 SRI
Appendix B Cable Pinouts Standard Cable Pinouts (Primary Interface Cables) RS-232 1 2 3 4 5 6 7 8 9 10 CTS RTS RXD TXD 10 Position RJ Pin 1 (Scanner End) VCC IN GND IBM Port 5B/9B/17 1 2 3 4 5 6 7 8 9 10 10 Position RJ DATA DATA + Pin 1 (Scanner End) VCC_IN GND Product Reference Guide B-1
Cable Pinouts IBM USB 1 2 3 4 5 6 7 8 9 10 10 Position RJ DD+ Pin 1 (Scanner End) VIN GND USB & USB Keyboard 1 2 3 4 5 6 7 8 9 10 10 Position RJ DD+ Pin 1 (Scanner End) VBUS_VIN GND Wand Emulation 1 2 3 WAND~ 4 5 6 7 8 VCC_IN 9 GND 10 B-2 10 Position RJ Pin 1 (Scanner End) PowerScan® 7000 SRI
Standard Cable Pinouts (Primary Interface Cables) Keyboard Wedge 1 2 3 4 5 6 7 8 9 10 KB_DATA AT_CLK KB_CLK AT_DATA 10 Position RJ Pin 1 (Scanner End) VCC_IN GND Product Reference Guide B-3
Cable Pinouts NOTES B-4 PowerScan® 7000 SRI
Appendix C Alpha-Numeric Pad A B C D E F Product Reference Guide C-1
Alpha-Numeric Pad 1 2 3 4 5 6 7 8 9 0 C-2 PowerScan® 7000 SRI
Appendix D Default Settings Standard Feature Defaults The table immediately below lists the default settings for the standard RS-232 interface. Feature Std. RS-232 Setting Double Read Timeout .
Feature Std. RS-232 Setting Interface-Specific Exceptions Page # Interface as required IBM Default: IBM Port 9B Interface Selection RS-232 Std.
Standard Feature Defaults Feature Std.
Std.
Standard Feature Defaults Feature Std.
Std.
Standard Feature Defaults Std.
Feature Std.
Standard Feature Defaults Feature Std.
Feature Std.
Standard Feature Defaults Feature Std.
Feature Std.
Standard Feature Defaults Feature Std.
Feature Std 2 of 5 Length 2 Std.
Appendix E Keyboard Function Key Mappings Keyboard Model Cross Reference Table E-2 summarizes the keyboard models, their defined protocol, scancode set, and some unique features. The remaining tables in this chapter provide the function key maps associated with each of the scancode sets.
Table E-1. Keyboard Model Cross Reference Model Type PC/XT Foreign ALT Mode AT; PS/2 25-286; PS/2 30-286; PS/2 50, 50Z; PS/2 60,70,80,90,95 Foreign ALT Mode PS/2 25 and 30 Foreign ALT Mode PC/XT U.S. Mode AT; PS/2 25-286; PS/2 30-286; PS/2 50, 50Z; PS/2 60,70,80,90,95 U.S. Mode + specific country support PS/2 25 and 30 U.S. Mode IBM 3xxx Terminals (122-key keyboard) IBM 3xxx Terminals (102-key keyboard) PS55 5530T with JAPANESE DOS (TDOS) NEC 9801 E-2 Wedge A Func.
Keyboard Model Cross Reference Table E-2.
Table E-3.
Keyboard Model Cross Reference Table E-4.
Table E-5.
Keyboard Model Cross Reference Table E-6.
Table E-7.
Keyboard Model Cross Reference Table E-8.
NOTES E-10 PowerScan® 7000 SRI
Appendix F Host Commands Accepting RS-232 Commands The scanner responds to the following RS-232 commands: COMMAND ASCII HEX COMMENT Enable Scanner E 0x45 Disable Scanner D 0x44 Reset Scanner R 0x52 Not On File Indication F 0x46 Long series of beeps Beep Good Read Tone B 0x42 Beeps if Good Read Beep is enabled Force Good Read Tone 0x01 Beeps regardless of beep setting Bel 0x07 Force Good Read Tone Identification request i 0x69 Returns long responsea Health request h 0x68 Re
NOTES F-2 PowerScan® 7000 SRI
Appendix G Sample Symbols UPC-A 0 123456 7890 Interleaved 2 of 5 Code 128 1234567890 Code 128 Code 39 EAN-13 9 780330 290951 Codabar BC321 Code 93 A13579B 123456-9$ Code 2 of 5 123456 Product Reference Guide G-1
Sample Symbols PDF417 Car Registration RSS-14 (01)00123456789012 RSS Expanded 0100123456789050 G-2 PowerScan® 7000 SRI
Appendix H Beeper/LED Indications LED and Beep Indications INDICATION LED BEEPER The scanner is in the process of powering-up. N/A Scanner beeps once for one second at highest frequency and volume upon power-up. Good Read Beep A label has been successfully scanned by the scanner.
INDICATION DESCRIPTION LED Scanner Active Mode The scanner is active and ready to scan. The LED is lit steadily. (ONLY applicable if LED Idle State is "On Dim") Scanner Disabled The scanner has been disabled by the host. The LED blinks continuously 100mS on / 900 mS off BEEPER N/A N/A Programming Mode - The following indications ONLY occur when the scanner is in Programming Mode. Label Programming Mode Entry A valid programming label has been scanned.
LED and Beep Indications INDICATION DESCRIPTION Label Programming Mode Acceptance of Programming Configuration option(s) have been successfully programmed via labels and the scanner has exited Programming Mode. Product Reference Guide LED BEEPER N/A Scanner sounds three beeps at current frequency volume for 80 mS on / 20 mS off.
Error Codes Upon startup, if the scanner sounds a long tone, this means the scanner has not passed its automatic Selftest and has entered FRU1 isolation mode. If the scanner is reset, the sequence will be repeated. The following table describes the LED flashes/beep codes associated with an error found.
Error Codes NOTES Product Reference Guide H-5
NOTES H-6 PowerScan® 7000BT SRI
ASCII Chart ASCII Char. Hex No. ASCII Char. Hex No. ASCII Char. Hex No. ASCII Char. Hex No. NUL SOH STX ETX EOT ENQ ACK BEL BS HT LF VT FF CR SO SI DLE DC1 DC2 DC3 DC4 NAK SYN ETB CAN EM SUB ESC FS GS RS US 00 01 02 03 04 05 06 07 08 09 0A 0B 0C 0D 0E 0F 10 11 12 13 14 15 16 17 18 19 1A 1B 1C 1D 1E 1F SP ! “ # $ % & ’ ( ) * + , .
Australia Japan Datalogic Scanning Pty Ltd Datalogic Scanning KK North Ryde, Australia Shinagawa, Tokyo, Japan Telephone: [61] (2) 9870 3200 Telephone: 81 (0)3 3491 6761 Fax: [61] (2) 9878 8688 Fax: 81 (0)3 3491 6656 France and Benelux Latin America Datalogic Scanning Sarl Datalogic Scanning, Inc LES ULIS Cedex, France Miami, Florida, USA Telephone: [33].01.64.86.71.00 Telephone: (305) 591-3222 Fax: [33].01.64 46.72.