User manual
Installation and Setup
2-2 ProMaster 2500 User Manual
Before You Install
Before you install the ProMaster 2500, confirm that you have the
following items ready in the area where it will be operating:
• Room temperature between 50 and 100 degrees F (10 and 38
degrees C).
• Clean, dry, oil-free factory air supply capable of supplying a constant
volume of 1.0 CFM (cubic feet/minute) at 80 PSI (pounds/square
inch) (5.5 Bar at 0.0283 CMM).
• Air filter (rated at 10 microns) and regulator located between your
compressor and the 2500.
• An air line from your filter/regulator with either a female quick
connect or female 1/4-inch NPT adapter to connect to the air line
supplied with the 2500. The 1/4-inch line provided with the system is
approximately 6 feet long. To prevent condensation in the lines, their
combined length should not exceed 10 feet.
Unpacking and Inspecting the 2500
This section describes how to remove the 2500 from its shipping crate and
lists the components that are shipped with the 2500. Follow the steps
below to remove the 2500 from its wooden shipping crate.
Note: Examine the wooden shipping crate for signs of damage before you begin
to unpack the 2500. Two indicators are attached to the outside of the
shipping crate: the drop indicator tells you if the crate was dropped, and
the tilt indicator tells you if the crate was placed on its side. If either is
positive, or if the crate is damaged in any way, contact your shipping
agent and Data I/O Customer Support (refer to the Preface).
Figure 2-1
Connecting the Air Line from the External Source to the 2500
INPUT AIR
LINE (6 ft.)
ADAPTER
MALE QUICK CONNECT
(With female 1/4 NPT)
FEMALE QUICK CONNECT
(Knurled Sleeve)
SOURCE AIR
LINE (4 ft. max)
FILTER/REGULATOR
TO EXTERNAL
AIR SOURCE
10 FT MAX
SUPPLIED BY CUSTOMER DELIVERED WITH PROMASTER
1761-1