User manual
Local Mode
F-12 ProMaster 2500 User Manual
The 2500 displays:
Press
1
. The 2500 displays:
The
C
in the displays represents the category signal sent by TaskLink (or
any computer remote control driver program) to indicate the result of the
last programming operation. In the example above, a
C1
signal from
TaskLink indicates that the device passed.
BIN 1
represents output tube
holder 1 and
BIN 2
represents output tube holder 2. When TaskLink send
a C1 signal, the 2500’s beam takes the device to the labeler (if the Task has
label selected as part of the current process) and then places that device in
output track 1.
For a device that has had a programming error, TaskLink might be
configured to send a C2 to the 2500. The 2500 would not label the failed
device and would send it to output bin 2, which is output tube 2.
This is the default setting for the binning control setup. Notice that in the
second line for all failures, there is no
X
after
Label
in the second line of
the display above. If labeling has been selected in the Task, this
unselected option instructs the 2500 not to label a device that has failed. If
you find that the 2500 is labeling failed devices, check to see that someone
has not inadvertently placed an X after Label in the second line. The 2500
will not label passing devices with one label text and failing devices with
different label text.
Control Functions
This command is used in conjunction with the Program/Test Only and
Program/Test and Label commands under the Operations menu. These
commands are used by Data I/O service personnel only and are not
usable under normal operation.
* * * HANDLER SETUP MENU * * *
1 - BINNING
2 - CONTROL FUNCTIONS
3 - SETUP MENU
BIN 1=C1-X C2- C3- C4- C5- LABEL-X
BIN 2=C1- C2-X C3-X C4-X C5-X LABEL-