Preventive Maintenance
5-42 ProMaster 2500 User Manual
Cycle Parts
This option inserts devices into the programming module without
programming them. It differs from the diagnostic tests because it allows
you to assess the following system operations under operating conditions
without having to program devices:
• Beam movement
• Vacuum generator functionality
• High and low air pressure
• Applying labels to devices
To select Cycle Parts, press
4
from the Diagnostic menu. The 2500
displays:
To Cycle Only
If you do not want to label devices, press
N
at the
DO YOU WANT TO
LABEL?
prompt. The 2500 displays:
Use the arrow keys to select an existing device or describe a custom
device, and press
ENTER
. Only eight selections appear on the display at
once; use
↑
and
↓
to scroll to the selections not visible on the display.
The 2500 displays:
Use the 2500’s arrow keys to move the pin 1 icon to correspond to the
location of pin 1 on the device in the input tube, and press
ENTER
. The
2500 displays:
Use the arrow keys to select the pin 1 position of the device as it goes into
the receiving tube. Press
ENTER
when you are satisfied with your
selection.
The next display prompts you to enter the number of parts per tube. The
default value appears in parentheses.
DO YOU WANT TO LABEL? (Y OR N)
CYCLE PARTS
CUSTOM PLCC 20 PLCC 28 PLCC 32
PLCC 44 PLCC 52 PLCC 68 PLCC 84
MOVE TO SELECTION THEN PRESS ENTER
1 <------- TEMPORARY FILE
SELECT POSITION OF PIN 1
AS IT COMES OUT OF THE TUBE.
USE ARROW KEYS THEN PRESS ENTER.
<------- TEMPORARY FILE
SELECT POSITION OF PIN 1
AS IT WILL GO INTO THE TUBE.
1 USE ARROW KEYS THEN PRESS ENTER.
CYCLE PARTS
ENTER NUMBER OF PARTS PER TUBE (13): _