User's Guide

Table Of Contents
Section 4: AlarmView
TM
Programmer Overview
Part #880-0325-01 Rev. A 33
! When you save edits to cap codes, entries are automatically
resorted in numeric order. The cap codes you changed or
entered will move in the display to appear in their new place in
the sort, not necessarily in the field where you made the
changes. To view an edit you’ve made, look for the cap code in
its correct place in the numeric sort.
! Lists of all cap codes in the AlarmView System are retained
permanently in the programmer. The values in these lists are the
values that will be selectable when programming the transmitter.
These lists should be set up during installation of the AlarmView
System and should not be changed unless pagers need to be
added or replaced.
Viewing and Changing Bed and Pager
Assignments
One bed is assigned to each transmitter. You must change the
bed number assignment if you move the transmitter to a different
bed or room.
One, two, or all pagers may be assigned to each transmitter. You
can change pager cap code assignments if caregivers or pagers
change. If you assign All Pagers, the transmitter will transmit to
all pagers on the AlarmView System.
u Do not exchange transmitters or pagers from different systems.
Because transmitters and pagers within a system operate on a
single, defined frequency, interchanged transmitters or pagers
will not correctly transmit or receive.