Part #880-0326-01 REV. 10/4/2000 Appendix B-7
10. Repeat steps #8 and #9 for each transmitter in the care unit.
If one of the pagers does not receive the page, the transmitter
may not be working or may be incorrectly programmed.
11. Turn off all transmitter, pagers, and the programmer.
Final Steps
The purpose of this activity is to ensure that all transmitters are
correctly transmitting to all the pagers.
1. Plug in each AlarmView transmitter to AC power for 17
hours to ensure the batteries are fully charged before
placing the AlarmView transmitters into service.
Select the location the clinical staff will store the AlarmView programmer
when not in use. Plug in the AC power cable at that location and instruct
the clinical staff to plug in the programmer when it is not in use.