User Manual
AlarmView™ System Administrator’s Manual
Part #880-0326-01 REV. 10/4/2000 Appendix B-6
2. With the programmer, use the AlarmView SysMgr program and
go to the ASSIGNMENTS screen.
3. Select any BED NUMBER and select ALL PAGERS. Press the
SEND button to program the transmitter.
4. Verify that all the pagers received a "Programming OK / All
Pagers" message.
If one of the pagers does not receive a page, that pager’s
common cap code may be incorrect. If none of the pagers
receive the page, either the programmer or the pagers may have
an incorrect common cap code.
The following steps verify that the Programmer and the Pagers have
the correct individual cap code.
5. With the programmer, select any BED NUMBER and select an
individual pager. Press the SEND button to program the
transmitter.
6. Verify that the selected pager received a "Programming OK /
xxxx" message where ‘xxxx’ corresponds to the last 4 digits of
the pager’s serial number.
7. Repeat steps #5 and #6 for each pager in the care unit.
If one of the pagers does not receive a page, either the
programmer has the wrong cap code entered or that pager’s
individual cap code may be incorrect.
The following steps verify that each Transmitter is working properly.
8. For each transmitter in the system, program it to any of the
pagers and bed numbers.
9. Verify that the selected pager received a "Programming OK /
xxxx" message where ‘xxxx’ corresponds to the last 4 digits of
the pager’s serial number.