User Manual

AlarmView™ System Administrator’s Manual
Part #880-0326-01 REV. 10/4/2000 Appendix B-1
INITIAL INSTALLATION PROCEDURE
This procedure should be performed upon initial installation of an
AlarmView System into a care unit.
Setup
1. Unpack the equipment to be installed in the care unit.
2. Determine the furthest limits of the care unit based on physical
distances.
Frequency Verification
The object of this activity is to use the transmitter’s carrier detect ability
to determine if other sources of potential RF noise exist on the selected
frequency.
1. Program the AlarmView transmitters to the desired frequency
assigned to the care unit. (See Section 4: Programming.)
2. Program the appropriate pager baud rate into the AlarmView
transmitters. (See Section 4: Programming.)
3. Program the All Pagers cap code into the AlarmView transmitter
if different than 0000031. (See Section 4: Programming.)
4. Program Survey Page mode for 150 pages at an interval of 120
seconds. (See Section 5: Testing and Periodic Maintenance.)
5. Place several AlarmView transmitters throughout the care unit.
6. One by one, perform steps #7 and #8.
7. Turn on a transmitter and activate Survey Paging. (See Section
5: Testing and Periodic Maintenance.)
8. Observe the Transmitting Indicator LED.