User Manual
AlarmView™ System Administrator’s Manual
880-0326-01 REV. A Page 60
Please be sure the
transmitter is turned
on, the IR port is
pointed directly at the
transmitter, at a
distance less than 4
ft. (1.2 m) and try
again.
Programmer infrared
port not aligned with
transmitter infrared
port or too far away
from transmitter.
Transmitter turned
off.
Programmer batteries
are low.
Transmitter defective.
Programmer
defective.
Re-align ports, ensure
the distance between
the transmitter and the
programmer is less
than 4' (1.2 m) and try
again.
Turn on transmitter.
Recharge programmer
batteries.
Contact Data Critical
regarding repair and/or
replacement of
transmitter or
programmer.
Programming Failed
Meaning:
Programmer was
unable to send
programming
changes to
transmitter.
Programmer infrared
port not aligned with
transmitter infrared
port or too far away
from transmitter.
Interference or
ambient light noise.
Programmer batteries
are low.
Transmitter defective.
Programmer
defective.
Re-align ports, ensure
the distance between
the transmitter and the
programmer is less
than 4' (1.2 m) and try
again.
Recharge programmer
batteries.
Contact Data Critical
regarding repair and/or
replacement of
transmitter or
programmer.