User Manual

AlarmView™ System Administrator’s Manual
880-0326-01 REV. A Page 58
Battery and Power Troubleshooting
Message Possible Cause Corrective Action
Low Tx
Battery/Connect AC
Power to Transmitter
Meaning:
Transmitter’s battery
is running low
(connect transmitter
to AC power as soon
as possible)
Transmitter is not
connected to AC
power.
Connect transmitter to
power.
Tx Battery
Low/Shutting Down
Meaning: Transmitter
operating on battery
power has depleted
battery charge and
will shut down
immediately.
Transmitter is not
connected to AC
power.
Connect transmitter to
power.
Comm
Failure/Retrying
Meaning:
Communication has
been lost between
transmitter and
monitor, the
transmitter is
attempting to
reestablish
communication.
Communications
between transmitter
and monitor have
failed.
Monitor is turned off
or operating on
battery power and
transmitter is still on.
Check serial cable and
cable connections
between transmitter
and monitor.
Reconnect or replace
cable as required.
Connect monitor to AC
power.
Check monitor
performance.