User Manual
AlarmView™ System Administrator’s Manual
880-0326-01 REV. A Page 55
Comm Failure
Retrying
Communication has
been lost between
transmitter and
monitor, the
transmitter is
attempting to
reestablish
communication.
Communications
between transmitter
and monitor have
failed.
Monitor is turned off or
operating on battery
power and transmitter
is still on.
Check serial cable
and cable connections
between transmitter
and monitor.
Reconnect or replace
cable as required.
Connect monitor to
AC power.
Check monitor
performance.
Comm
Failure/Shutting Down
Transmitter has not
been able to
communicate with the
monitor for about an
hour and is shutting
down.
Communications
between transmitter
and monitor have
failed.
Monitor is turned off or
operating on battery
power and transmitter
is still on.
Check serial cable
and cable connections
between transmitter
and monitor.
Reconnect or replace
cable as required.
Connect monitor to
AC power.
Check monitor
performance.