5. The transmitter will transmit a series of consecutively numbered
pages (1, 2, 3, 4, etc.). Assigned pagers will receive the following
message for each numbered page:
Bed: 1234
Survey Page #xxx*
* XXX is the number of the Survey Page in the sequence.
6. Walk slowly around the unit, observing the sequentially
numbered pages. Pay special attention to elevator banks,
columns, metal doors, double doors, and areas with plaster
construction. Areas of no coverage or out of range conditions will
be identified by missing Survey Page numbers in the sequence.
A corrupt, partial or incomplete page is the same as a missing
page.
7. To stop the transmitter before all Survey Pages are sent, press
and release the On/Off button on the transmitter once.
If any areas of no coverage are defined, communicate their location to
clinical staff.
AlarmView is intended to cover most care units, not entire floors or entire
facilities. Typical range for the AlarmView System with no intervening
structures is 150 feet.
Contact Data Critical Corporation if areas of no coverage present a
problem in functioning of care unit.