To send a Test Page:
1. Press the On/Off button to turn the transmitter on. Check to
ensure that the transmitter is communicating with the
monitor (both AC Indicator and Communications Indicator
on transmitter are green).
2. Turn on the pager(s) assigned to the transmitter.
3. To send a Test Page, press and release the On/Off button
again.
All pagers assigned to the transmitter will display the
following page (if a patient is not connected to the sensor,
dashes will be displayed for SpO2 and pulse rate):
Bed: 1234
Test Page
SPO2: 83%
PR: 70 bpm*
* For the Nellcor N-3000 monitor, the last line of the test page
will display heart rate (HR).
Transmitter Range Test (Survey Paging)
Survey Paging is used to determine the area of coverage of your system.
A Range Test (Survey Paging) will be performed at the time of initial
system installation. Survey Paging should be conducted prior to the first
use of a transmitter added to a system, or when a transmitter has been
moved to a new care unit.
When the transmitter is placed in Test Range mode, it will send frequent
pages from the transmitter to the assigned pager(s). Frequency of sent
pages and the total number of Survey Pages can be set according to the