880-0326-01 REV. A Page 47
Indicators will flash on and off while transmitter performs a
Power-On Self Test (POST) to check its function.
Mode LED is solid green when transmitter has established
communication with the monitor.
Tx LED will flash green, and the transmitter will send a Comm
OK Page to assigned pager(s) to confirm communication
between transmitter and monitor. The message will display:
Bed: 1234
Comm OK
All Pagers*
*Last line All Pagers with be replaced with “Pager 1234” if one
pager is assigned to transmitter, and “Pager 1234 5678.” The 4
digits correspond to the last 4 digits of the assigned pager(s) cap
code.
3. Verify that the page is sent to the assigned pager(s) and that
the bed number is correct.
The transmitter will now send pages to the assigned pager(s)
whenever it detects alarms identified by the monitor.
Test Page
You should test the AlarmView System setup by conducting a
Test Page when any changes affecting the system have
occurred. To perform a Test Page, the transmitter must be on
and communicating with the monitor.