User Manual
AlarmView™ System Administrator’s Manual
880-0326-01 REV. A Page 45
Section 5: Testing and Periodic
Maintenance
All tests described in this section can be performed without
disassembling the units in the AlarmView System. Tests should be
conducted prior to initial installation, when adding a new unit to the
system, and before returning equipment to service after repairs or routine
maintenance.
If any component of the system fails to perform as specified, do not
return it to use. Contact Data Critical Corporation regarding repair or
replacement of the equipment.
Testing System Safety
AlarmView safety tests meet the standards of, and are to be performed
in accordance with, CAN/CSA-C22.2 No. 601.1 and UL 2601-1, Medical
Electrical Equipment – Part 1: General Requirements for Safety, for
instruments classified as Class 1 and Type B.
!!
!!
!!
Equipment Needed is a Safety Analyzer that must meet ,
CAN/CSA-C22.2 No. 601.1 and UL 2601-1 specifications.
Earth Leakage Current
This test is in compliance with UL 2601-1 (earth leakage
current). The applied voltage is 120 volts, 60 Hz for UL 2601-1
and CSA 601.1. Make all measurements with power switch in
both the On and Off positions.
1. Connect the AlarmView transmitter power supply AC plug to
the electrical analyzer as recommended by the analyzer
operating instructions.
2. Connect the AlarmView transmitter power supply DC plug to
the transmitter DC input connector.