Turning Transmitter Off
The transmitter can be turned off only when it is not
communicating with the monitor. To turn the transmitter off:
1. Turn monitor off or disconnect interface cable from
transmitter or monitor. Mode LED will be amber.
2. Press On/Off button. Mode LED will be Off.
!!
!!
!!
If you disable communications between the transmitter and the
monitor and do not turn off the transmitter, the transmitter will
send Communication Failure pages to the assigned pagers at
periodic intervals for up to one hour.
Operating Transmitter On Battery Power
The transmitter has an internal, rechargeable backup battery. It can
operate on battery power for approximately 45 minutes (under normal
operation) if it becomes disconnected from AC power.
The battery will automatically recharge when connected to AC power.
Solid green AC LED indicates that the transmitter is connected to power,
and that the internal battery is recharging. A fully discharged battery
takes approximately 17 hours to return to full charge. The transmitter can
be used while it is recharging.