User Manual
AlarmView™ System Administrator’s Manual
880-0326-01 REV. A Page 49
procedure in Section 4: Programming, “Changing Survey Page
Configuration” of this manual.
As Survey Pages are sent, personnel with assigned pager(s) will walk
around the coverage area of the care unit to ensure that all pages are
received.
Because transmitters and pagers in one system operate on a single,
defined frequency, you should NOT exchange pagers or
transmitters from different systems.
To perform Survey Paging for a transmitter, transmitters should be
turned on, but not communicating with monitors. You will need at least
one transmitter and one pager.
To conduct a transmitter Range Test (Survey Paging):
1. Ensure that transmitter is connected to power and turned on, but
not communicating with a monitor.
2. On the programmer, from the SysMgr main screen, tap Service
to enter service program. The Controls tab will be displayed.
3. To start Survey Paging, aim programmer’s infrared port directly
at the port on the front of transmitter and tap the Test Range
button
. The programmer will display a “Survey Command
Sent” status message and Survey Paging will begin.
4. Turn on the pager(s) assigned to transmitter. Qualified personnel
should be ready to walk the dimensions of the care unit. If you
use more than one pager, keep them at least 12” apart to avoid
interference.