User's Guide
Environment - Data Critical Corporation does not assume responsibility
for damage to the equipment caused by improperly vented cabinets,
improper or faulty power, or insufficient wall strength to support
equipment mounted on such walls.
Explosion Hazard - Do not use this equipment in the presence of
flammable anesthetics, skin cleaning agents, and disinfectants. Great
care must be exercised when the device is used in an oxygen-enriched
atmosphere. The atmosphere is considered to be oxygen enriched when
the room air contains more than 25% of oxygen or nitrous oxide.
Home Use - AlarmViewâ„¢ System is not intended for home use.
Power - Convenience outlets and all others that may be used for the
patient must be on the hospital emergency circuit so that they are
energized at all times.
Site Surveys - Each institution is responsible for surveying existing
equipment and ensuring no conflict exists when considering the
introduction of additional radiating devices.
Transmission Bands - Each institution is responsible for maintaining
continued surveillance of transmission bands being used in the facility to
prevent co-occupation of bands and cross-band interference.