User Manual
ALARMS
PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft 123
All information contained herein is subject to the rights and restrictions on the title page.
Table 15. Bedside device technical alarms
Alarm Label Alarm Triggering Condition
SYS ERR system error bedside monitor device equipment failure, printer failure, or
printer paper error
ECG ERR ECG error bedside ECG equipment failure
BP ERR BP error non-invasive equipment failure; uncalibrated; no pulse
detected
RESP ERR respiration error no probe detected
P
1
ERR P
1
error P
1
not zeroed
P
2
ERR P
2
error P
2
not zeroed
P
3
ERR P
3
error P
3
not zeroed
P
4
ERR P
4
error P
4
not zeroed
P
5
ERR P
5
error P
5
not zeroed
P
6
ERR P
6
error P
6
not zeroed
SpO
2
ERR SpO
2
error equipment SpO
2
failure, uncalibrated
CO
2
ERR CO
2
error equipment CO
2
failure, uncalibrated
T
1
ERR T
1
error equipment T
1
failure, uncalibrated
T
2
ERR T
2
error equipment T
2
failure, uncalibrated
NO ECG no ECG ECG not detected
NO BP no BP BP not detected
NO P
1
no P
1
P
1
not detected
NO P
2
no P
2
P
2
not detected
NO P
3
no P
3
P
3
not detected
NO P
4
no P
4
P
4
not detected
NO P
5
no P
5
P
5
not detected
NO P
6
no P
6
P
6
not detected
NO SpO
2
no pulse oximetry SpO
2
not detected
NO CO
2
no CO
2
CO
2
not detected
NO T
1
no T
1
temperature probe not detected
NO T
2
no T
2
temperature probe not detected
NO RESP no respiration respiration not detected
?? SOURCE source data received does not match configured device type
NURSE nurse call nurse call button on transceiver was pressed
LOW BATT low transceiver battery low battery on the V-Link/V-Link II transceiver
DEV LOWBATT low device battery low battery level in the bedside monitor
MUSCLE muscle artifact muscle artifact detected
WRONG ID wrong ID incompatible transceiver ID
EDIT edit number of history events reached the edit limit
LEAD OFF lead off ECG electrode disconnected or loose
NO SIGNAL no signal no signal received from transceiver